37
PMA1000/PMA1000L | 1.00.00
www.mc-techgroup.com
When you tap on the edit button close to the measuring range the highlighted scroll bar
opens. The active edit button changes to a green check mark. Please scroll through the
predefined measuring ranges by swiping vertically.
The selected measuring range needs to be displayed in the gray frame in the middle of
the scroll bar. Please tap on the green check mark to confirm your selection.
2
1
Fig. 34: Highlighted scroll bar to select measuring range
1
Scroll bar to select measuring range
2
Active edit button changes into a check mark
You will find a more detailed description about the measuring range selection on page 36
chapter ‘M4/S2 - Settings menu/ parameters’ .
The parameter list shows the current (‘real’), as well as the factory set gradient and offset
parameters. The real gradient and offset can deviate from the factory settings as long as
the values are staying in the stated range. Is the current gradient or offset higher or lower
than the permitted range, the indicator below the edit button turns from green to red.
2
1
Fig. 35: Parameter list
1
Edit button to open parameter table
2
Parameter list