46
Battery capacity measured according to IEC 60601-2-4, clause
102.3.2 at room temperature. Capacity may be diminished at
operating temperature extremes, or when the available battery
charge is used in multiple power ON/OFF cycles.
7.5
Electrodes
Note
Do not use defibrillation pads not approved by M&B. Use of
unauthorized defibrillation pads may cause the device to
operate improperly and provide false measurements. Follow all
labeling instructions on the defibrillation pads.
7.6
Guidance and manufacturer's declaration
7.6.1
Electromagnetic emissions (IEC 60601-1-2 Table 201)
The AED7000 is intended for use in the electromagnetic environment specified
below. The customer or the user of the AED7000 should assure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic environment –
guidance
RF emissions
CISPR 11
Group 1
TheAED7000 uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not
applicable