14
Follow instructions in the Users Manual.
2.5
Battery
2.5.1
Battery Care
Caution
Use only M&B batteries
Use only M&B Part No. CR123A-4x 2
non-rechargeable lithium batteries. Use of any
other battery can damage the AED7000.
Caution
Always verify remaining capacity of a
non-rechargeable battery after use
Check capacity and change if battery is low.
Caution
Replace battery at 5 years
Battery replacement at 5 years is recommended
due to degradation of the battery chemistry.
Periodic maintenance and testing is highly
recommended to ensure proper battery
performance.
2.6
Care and Storage
2.6.1
Cleaning and Sterilization
Caution
Clean and maintain the AED7000 according to
instructions. See Chapter 4, Maintaining the
AED7000 Automatic External Defibrillator.
Do not clean the AED7000 with alcohol,
ketone, or any flammable agent. Do not
autoclave the AED7000 or attempt to sterilize
the AED7000 or any of its accessories.