49
MAINTENANCE INTERVALS
Every 2 years and after every emergency lowering
Replace the two 9-V block batteries
Annually
Inspection and maintenance
As required
Lubricate mechanical components
Replace worn components
if a defect occurs.
DELIVERY AND ASSEMBLY
Betten Malsch GmbH care beds are generally delivered fully as-
sembled, or are assembled on site by company technicians or
contractual partners.
Check the delivered bed against documentation for completeness
and conformity.
Any defects or damage must be pointed out to the freight com-
pany immediately and noted on the delivery document.
Signing of the delivery documents by both parties is obligatory
before commissioning.
If necessary, e.g. for maintenance, simple assembly procedures can
also be performed by professional authorised persons.
After maintenance and servicing work has been
completed, check the functionality of electrical sys-
tems.
The manufacturer off ers technician training for
maintenance and servicing work on their care beds.
A certifi cate obtained as part of this training au-
thorises the holder to carry out technical work in-
dependently on the beds.