10
|
Observe safety distances to walls, window ledges and other
furnishings when using the care bed in an resident’s room.
The safety distances depend on the design and model of the
care bed and are based on the height adjustment and the
tilting motions of the bed. The minimum distance is 30 mm.
Improper use of the bed may cause hazards. Examples of
improper use include:
Unauthorised activation of the electrical functions
Use of the bed by persons with a body weight of less
than 40 kg or a BMI of less than 17, or a height of less
than 146 cm
p. 7 Specifi c function
Moving the bed by pulling on the mains cable or side rails
More than one person adjusting the bed at the same time
Activation of the functions by the resident without prior
instruction
Pulling the mains cable to disconnect it from the power
supply
Moving the bed on sloping or unsurfaced ground
In line with the standard DIN EN 60601-2-52:2016, when
choosing a mattress, it is important that there is a minimum
distance of 22 cm between the top of the sleeping surface
and the top of the side rails in its fully extended position. The
mattress used must meet the applicable safety standards.
The constant presence of liquid in the area of the motor
must be avoided (e.g. incontinence).
For safety reasons, the handle on the trapeze bar must be
replaced completely every 5 years.
Servicing and repairs on electrical components must be car-
ried out by specially trained staff and only original replace-
ment parts from the manufacturer must be used.
The care bed is not suitable for extended operation be-
yond a working cycle of two minutes. If the mains adapter
is overloaded or if it overheats, it will shut off automatically.
Further operation is possible only after a 30-minute cool-
down phase. Observe the drive manufacturer’s notes on
the rating plate!
It is essential to avoid obstructing any part of the bed mech-
anism as this can lead to damage or complete disabling of
the drive system due to overheating.
Likewise, the safe working load must not be exceeded.
If an immobile resident remains in the same position for a
long period without the use of additional positioning aids,
this can lead to pressure sores. The manufacturer of the
care bed is not liable for this in any way.
Electrically operated care beds are active medical products
and must be maintained according to Article 7 of the Med-
ical Devices Operator Ordinance (MedProd-BetrV.) These
maintenance measures must be carried out regularly (at