Malmet Bedpan / Urinal Bottle Washer Disinfector (ES-D)
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 5
Page
32
June 23
Fault 10
Disinfection temperature not maintained, resulting in cycle time out
Cause
•
Steam leak
•
Final rinse solenoid stuck open
•
Faulty temperature sensor hot water
tank
•
Blown element leg
•
Steam condensate leaking cold water into chamber
•
Faulty temperature sensor - chamber
Fault 12
Hot tank short temperature sensor
Cause
•
Faulty temperature sensor steam generator
Fault 13
Hot tank open circuit temperature sensor
Cause
•
Faulty temperature sensor steam generator
Fault 14
Disinfection chamber short circuit temperature sensor
Cause
•
Faulty temperature sensor - chamber
Fault 15
Disinfection chamber open circuit temperature sensor
Cause
•
Faulty temperature sensor - chamber
Fault 30
dEt
Insufficient detergent last wash
Cause
•
Empty detergent bottle
•
Faulty detergent pump
•
Blocked detergent line
Fault 31
HEAt
Disinfection temperature not maintained on 2
nd
probe
Cause
•
Steam leak
•
Final rinse solenoid stuck open
•
Faulty temperature sensor hot water tank
•
Blown element leg
•
Steam condensate leaking cold water into chamber
•
Faulty temperature sensor
–
chamber
•
Faulty temperature sensor
–
chamber 2
Fault 32
CFSd
Unsupported fault
(cycle did not follow supported order)
Cause
•
Interrupted power
•
Processor glitch
Fault 33
bLOC
Chamber blockage present
–
flood sensors
(Sensors are detecting a build-up above the bottom of the door inside the
chamber)
Cause
•
Blocked trap
•
Blocked / backed up drain
•
Dirty sensors
Fault 34
dEtH
Extra detergent injected last wash
(the volume of detergent injected during the cycle was greater than necessary)
Cause
•
Faulty flow meter
•
Faulty detergent pump relay
•
Failed check valve
Fault 35
PobE
Temperature discrepancy
–
temperature difference >5˚C between probes
Cause
•
Faulty temperature sensor/s
•
Fluctuating temperatures
Fault 38
FPri
Failed priming
–
exceeded allowed priming attempts
Cause
•
Detergent bottle not changed
•
Detergent line split / leaking
•
Flow meter faulty
•
Check valve failed