4.0 Technical Specifications
4.1 Device Specifications
Capacity
(Single)
Blankets
105 Lt Models: 4 maximum load
210 Lt Models: 16 maximum load
260 Lt Models: 18 maximum load
420 Lt Models: 30 maximum load
Shelves
105 Lt Models: 2 adjustable
2 max load per shelf
210 Lt Models: 2 adjustable
8
max load per shelf
260 Lt Models: 3 adjustable
6 max load per shelf
420 Lt Models: 5 adjustable
6 max load per shelf
Electrical Rating
Volts
240Vac
Phase / Hz
1 ph / 50Hz
Amps
2.8 Amps
Watts
0.6 kW
Environment operating conditions
Temperature
Relative Humidity
+10
C to +25
C
+30% to +70%
Electrical Connection
IEC right angle Power Cord with
3 Pin Plug
10 Amp (into standard GPO 240
Volt)
Hot Air Fan
Continuous operation
Ball bearings 230V 50Hz
Elements
Circular Unfinned
Incoloy 840 sheathed 10mm
diameter tubular
Rated
240V 600 watt
Air cooling fan Control PCB
compartment
Continuous operation
220-240 VAC 50/60 Hz
Control PCB
Microprocessor Control Relay
PCB (SKCM325) Access via
compatible RS232
Temperature Control
Incremental 1°C
–
60°C
Factory set at 60°C
Over-temp Protection
Software
5°C above set point
Secondary Over-temp Protection
Mechanical
Auto reset Bi-Metallic Disc 80°C
Materials
Doors
Bi-parting, double glazed with
powder coated frames
Cabinet
304/4 and 430D stainless steel
Environment conditions Transport
and Storage
Fragile
Keep away from rain
Do not stack