Specifications

INOblender

Ü

  NO Set Range 

5 to 80 ppm (800 ppm cylinder)

   

2.5 to 40 ppm (400 ppm cylinder)

 

  Setting accuracy 

+/- 20 % indicated or 2 ppm, whichever is greater

 

  Gas flow range 

5 to 14 L/min of O

2

 

 O

2

 inlet connection 

DISS (male), 3.4 bar (50 psig, nominal)

 

  NO/N

2

 inlet pressure 

1.7 to 2.4 bar (25 to 35 psig)

 

  NO inlet  

Quick disconnect with shut-off (Female)

 

  NO/O

2

 outlet  

Barb fitting for oxygen supply tubing

 

  Max. Dimensions 

200 mm (W) X 120 mm (H) X 110 mm (D) DISS (clamp included)

 

   

200 mm (W) X 120 mm (H) X 131 mm (D) NIST (clamp included)

 

  Max. weight 

1.7 kg (clamp included)

Operating and Storage Conditions 

 

  Operating Temperature 

10 to 40°C

 

  Operating Humidity 

15% to 95% non-condensing

 

  Operating Barometric Pressure  57 to 110 kPa

 

  Storage Temperature 

-40 to + 70°C

 

  Storage Humidity 

15% to 98% non condensing

 

  Storage Barometric Pressure 

57 to 110 kPa

INOMAX Pressure Regulator

  Cylinder Valve Connector 

CGA 626  or ISO 5145 valve

 

  Inlet Pressure 

14 to 155 bar (200 to 2,248 psig)

 

  NO Outlet Pressure 

1.7 to 2.4 bar (25 to 35 psig) 

Manual Resuscitators

The INOblender is compatible with most types of manual resuscitation bags when used within the specifications of this 

device. The INOblender has been validated against the following manual resuscitation bags.* 

  * Airlife Adult Manual Resuscitator with reservoir tubing

 

  * Allegiance ½ L Pediatric Manual Resuscitator with reservoir tubing

Содержание INOblender 10026

Страница 1: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 Operation and Maintenance Manual English...

Страница 2: ...e faulty maintenance improper repair damage or alteration by anyone other than Mallinckrodt Representatives Outside the U S A and Canada check local laws for any restrictions that may apply Caution U...

Страница 3: ...roduction 4 Definitions and Abbreviations 4 Theory of Operation 5 Environmental Effects 6 2 Setup 7 3 Pre Use Checkout 9 4 Operation 12 5 Maintenance 15 Replacing Regulator Tips 16 Cylinder Leak Check...

Страница 4: ...Part No 20732 Rev 02 2014 02 2 Intentionally left blank...

Страница 5: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 1 General Information General Information...

Страница 6: ...Part No 20732 Rev 02 2014 02 1 General Information General Information...

Страница 7: ...y sensitive to NO or NO2 or who may be exposed to these agents for prolonged periods as a result of the use of this device should be aware that the INOblender does not scavenge the exhaust drug and th...

Страница 8: ...to be mixed into a user set flow of oxygen which is delivered to a patient The intended use for the INOblender is as a back up to a primary INOMAX delivery system or for short term attended use when a...

Страница 9: ...INOblender is also connected to a 2 bar 30 psig INOMAX 800 ppm or 400 ppm supply The INOMAX flow is controlled by a balance regulator in proportion to the oxygen pressure caused by the fixed restrict...

Страница 10: ...easurements as performed by Hess et al The NO and NO2 concentrations were measured using a chemiluminescence analyzer when 100 ppm of NO at 8 L min was delivered into a room with no scavenging being u...

Страница 11: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 2 Setup Setup...

Страница 12: ...Part No 20732 Rev 02 2014 02 2 Setup Setup...

Страница 13: ...pply to the O2 inlet fitting see part list for correct oxygen hose Note 3 5 bar 50 psig nominal 1 Make sure the NO dose setting dial is turned to zero 2 Attach a manual resuscitation bag to the INOble...

Страница 14: ...Part No 20732 Rev 02 2014 02 8 Intentionally left blank...

Страница 15: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...

Страница 16: ...Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...

Страница 17: ...nnot be traced replace the INOblender Delivery Confirmation and Purge 1 Set the INOblender to 40 ppm when using an 800 ppm cylinder 20 ppm when using a 400 ppm cylinder 2 Set the oxygen flow on the IN...

Страница 18: ...Part No 20732 Rev 02 2014 02 10 Intentionally left blank...

Страница 19: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 4 Operation Operation...

Страница 20: ...Part No 20732 Rev 02 2014 02 4 Operation Operation...

Страница 21: ...Part No 20732 Rev 02 2014 02 11 Intentionally left blank...

Страница 22: ...must be followed to help ensure NO2 is purged from the system before the manual resuscitator bag is connected to the patient The manual bag should be squeezed repeatedly during use to avoid NO2 from...

Страница 23: ...1 Turn the INOblender setting dial to the desired concentration 5 to 80 ppm for an 800 ppm cylinder and 2 5 to 40 ppm for a 400 ppm cylinder 2 Turn the O2 flowmeter to the desired flow rate 5 to 14 L...

Страница 24: ...Part No 20732 Rev 02 2014 02 14 Intentionally left blank...

Страница 25: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...

Страница 26: ...Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...

Страница 27: ...by Metrex Diisobutylphenooxyethoxyethyl dimethyl benzyl ammonium chloride 0 28 Isopropyl alcohol 17 2 Inert ingredients 82 52 Caution Apply cleaning agent to a cloth before application do not spray c...

Страница 28: ...SO see Figure 5 2 by pulling on the tip a and turning it counterclockwise b see Figure 5 1 3 Ensure the threads are clean on the regulator tip if required use a lint free cloth 4 Install the new tip F...

Страница 29: ...the O ring as they may damage the metal gland and cause a leak O ring Gland Replacing the O ring on the ISO 5145 INOMAX regulator fitting Note Depressurize the INOMAX regulator by using the purge port...

Страница 30: ...tightening the INOMAX regulator hand wheel a If cylinder valve is open close cylinder valve and tighten INOMAX regulator hand wheel b Open cylinder valve and reapply soapy water to points 1 and 2 c If...

Страница 31: ...1 INOMAX Regulator CGA 626 10006 INOMAX Regulator ISO 5145 10020 INOMAX Regulator Extension Hose 10014 Operation and Maintenance Manual 20732 O ring ISO 5145 INOMAX Regulator 80470 Tip CGA 626 INOMAX...

Страница 32: ...Part No 20732 Rev 02 2014 02 20 Intentionally left blank...

Страница 33: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 6 Product Specifications...

Страница 34: ...Part No 20732 Rev 02 2014 02 6 Product Specifications...

Страница 35: ...ting and Storage Conditions Operating Temperature 10 to 40 C Operating Humidity 15 to 95 non condensing Operating Barometric Pressure 57 to 110 kPa Storage Temperature 40 to 70 C Storage Humidity 15 t...

Страница 36: ...Part No 20732 Rev 02 2014 02 22 Intentionally left blank...

Страница 37: ...17 Part No 20732 Rev 02 2014 02 23 Intentionally left blank...

Страница 38: ...565 Europe European Authorized Representative MDSS GmbH Schiffgraben 41 30175 Hannover Germany Part No 20732 Rev 02 2014 02 Manufacturer North America Mallinckrodt Manufacturing LLC 6603 Femrite Driv...

Отзывы: