Part No. 20732  Rev-02

2014-02

4

Pneumatic Outlet

Rx ONLY

Prescription use only

Pressure

Product packaging is able to be

 

recycled

Protect from heat and radioactive sources

SN

Serial Number

Storage Temperature Limitation

Storage Humidity Limitation

This product is in compliance with the 

European Council Directive 93/42/EEC 

and its amendments concerning medical 

devices

Attention, consult accompanying 

documents!

Authorized Representative in the 

European Community

Catalog Number

Date of Manufacture

Fragile, handle with care

Keep Dry

Manufacturer

NO or O

2

 Gas Inlet

NO Gas Outlet

Pneumatic Inlet

Introduction

The INOblender allows users to select a concentration of INOMAX, (nitric oxide in a balance of nitrogen), to 

be mixed into a user set flow of oxygen which is delivered to a patient. The intended use for the INOblender 

is as a back up to a primary INOMAX delivery system; or for short term attended use when a primary delivery 

device cannot practicably be used. This intended use includes applications within a medical facility and 

transport outside of a medical facility. The INOblender is not intended for use as a primary INOMAX delivery 

system for long term use.

Definitions and abbreviations

N

2

Nitrogen

NO

INOMAX (nitric oxide) for inhalation

NO

2

Nitrogen dioxide

NO/N

2

Nitric oxide (NO) and nitrogen (N

2

) gas mixture

O

2

Oxygen

ppm

Parts Per Million

Set NO

The dose of INOMAX set by the user

v/v

Volume to volume

Note: This manual shows the Set NO displays associated with the 0-80 ppm range.

Symbols used in this manual or on the system

Symbols replace words on the equipment and/or in this manual. These symbols include:

Содержание INOblender 10026

Страница 1: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 Operation and Maintenance Manual English...

Страница 2: ...e faulty maintenance improper repair damage or alteration by anyone other than Mallinckrodt Representatives Outside the U S A and Canada check local laws for any restrictions that may apply Caution U...

Страница 3: ...roduction 4 Definitions and Abbreviations 4 Theory of Operation 5 Environmental Effects 6 2 Setup 7 3 Pre Use Checkout 9 4 Operation 12 5 Maintenance 15 Replacing Regulator Tips 16 Cylinder Leak Check...

Страница 4: ...Part No 20732 Rev 02 2014 02 2 Intentionally left blank...

Страница 5: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 1 General Information General Information...

Страница 6: ...Part No 20732 Rev 02 2014 02 1 General Information General Information...

Страница 7: ...y sensitive to NO or NO2 or who may be exposed to these agents for prolonged periods as a result of the use of this device should be aware that the INOblender does not scavenge the exhaust drug and th...

Страница 8: ...to be mixed into a user set flow of oxygen which is delivered to a patient The intended use for the INOblender is as a back up to a primary INOMAX delivery system or for short term attended use when a...

Страница 9: ...INOblender is also connected to a 2 bar 30 psig INOMAX 800 ppm or 400 ppm supply The INOMAX flow is controlled by a balance regulator in proportion to the oxygen pressure caused by the fixed restrict...

Страница 10: ...easurements as performed by Hess et al The NO and NO2 concentrations were measured using a chemiluminescence analyzer when 100 ppm of NO at 8 L min was delivered into a room with no scavenging being u...

Страница 11: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 2 Setup Setup...

Страница 12: ...Part No 20732 Rev 02 2014 02 2 Setup Setup...

Страница 13: ...pply to the O2 inlet fitting see part list for correct oxygen hose Note 3 5 bar 50 psig nominal 1 Make sure the NO dose setting dial is turned to zero 2 Attach a manual resuscitation bag to the INOble...

Страница 14: ...Part No 20732 Rev 02 2014 02 8 Intentionally left blank...

Страница 15: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...

Страница 16: ...Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...

Страница 17: ...nnot be traced replace the INOblender Delivery Confirmation and Purge 1 Set the INOblender to 40 ppm when using an 800 ppm cylinder 20 ppm when using a 400 ppm cylinder 2 Set the oxygen flow on the IN...

Страница 18: ...Part No 20732 Rev 02 2014 02 10 Intentionally left blank...

Страница 19: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 4 Operation Operation...

Страница 20: ...Part No 20732 Rev 02 2014 02 4 Operation Operation...

Страница 21: ...Part No 20732 Rev 02 2014 02 11 Intentionally left blank...

Страница 22: ...must be followed to help ensure NO2 is purged from the system before the manual resuscitator bag is connected to the patient The manual bag should be squeezed repeatedly during use to avoid NO2 from...

Страница 23: ...1 Turn the INOblender setting dial to the desired concentration 5 to 80 ppm for an 800 ppm cylinder and 2 5 to 40 ppm for a 400 ppm cylinder 2 Turn the O2 flowmeter to the desired flow rate 5 to 14 L...

Страница 24: ...Part No 20732 Rev 02 2014 02 14 Intentionally left blank...

Страница 25: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...

Страница 26: ...Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...

Страница 27: ...by Metrex Diisobutylphenooxyethoxyethyl dimethyl benzyl ammonium chloride 0 28 Isopropyl alcohol 17 2 Inert ingredients 82 52 Caution Apply cleaning agent to a cloth before application do not spray c...

Страница 28: ...SO see Figure 5 2 by pulling on the tip a and turning it counterclockwise b see Figure 5 1 3 Ensure the threads are clean on the regulator tip if required use a lint free cloth 4 Install the new tip F...

Страница 29: ...the O ring as they may damage the metal gland and cause a leak O ring Gland Replacing the O ring on the ISO 5145 INOMAX regulator fitting Note Depressurize the INOMAX regulator by using the purge port...

Страница 30: ...tightening the INOMAX regulator hand wheel a If cylinder valve is open close cylinder valve and tighten INOMAX regulator hand wheel b Open cylinder valve and reapply soapy water to points 1 and 2 c If...

Страница 31: ...1 INOMAX Regulator CGA 626 10006 INOMAX Regulator ISO 5145 10020 INOMAX Regulator Extension Hose 10014 Operation and Maintenance Manual 20732 O ring ISO 5145 INOMAX Regulator 80470 Tip CGA 626 INOMAX...

Страница 32: ...Part No 20732 Rev 02 2014 02 20 Intentionally left blank...

Страница 33: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 6 Product Specifications...

Страница 34: ...Part No 20732 Rev 02 2014 02 6 Product Specifications...

Страница 35: ...ting and Storage Conditions Operating Temperature 10 to 40 C Operating Humidity 15 to 95 non condensing Operating Barometric Pressure 57 to 110 kPa Storage Temperature 40 to 70 C Storage Humidity 15 t...

Страница 36: ...Part No 20732 Rev 02 2014 02 22 Intentionally left blank...

Страница 37: ...17 Part No 20732 Rev 02 2014 02 23 Intentionally left blank...

Страница 38: ...565 Europe European Authorized Representative MDSS GmbH Schiffgraben 41 30175 Hannover Germany Part No 20732 Rev 02 2014 02 Manufacturer North America Mallinckrodt Manufacturing LLC 6603 Femrite Driv...

Отзывы: