background image

13

3. Press 

 (ENTER/SAVE), LCD displays the picture shown 

in below:

4. Press 

 value-add “+" or 

 value-reduce "-" to set the 

test time.

5. After setting is complete, press 

 (ENTER/SAVE) to 

confirm & save the data measured.

6. Press 

 (ESC) to return back to the main page.

7.1.6. Display the data stored (LOG Display) ─ Function 4   

1. Press 

 (ON/OFF) button. 

(Main page)

2. Press 

 (value-add) "+" 4 times; LCD display the 

following pictures respectively:

3. Press 

 (ENTER/SAVE), LCD displays the picture shown 

below:

4. Press 

 (ENTER/SAVE), LCD displays the "DAR" and "PI" 

picture shown below:

Содержание MT585

Страница 1: ...INSTRUCTION MANUAL 15kV HIGH VOLTAGE INSULATION TESTER MT585...

Страница 2: ...2...

Страница 3: ...ge no 1 Safety Precautions 4 2 Overview 5 3 Features 5 4 Specifications 6 5 Connections 7 6 Instrument layout 8 7 Measuring procedure 9 8 Charge 16 9 Maintenance repair 16 10 Interface Connection and...

Страница 4: ...or use the tester Return your high voltage insulation tester to your nearest distributor for service 1 4 Your high voltage insulation tester has been designed with your safety in mind However no desi...

Страница 5: ...controlled 2 Lines x 16 characters large intelligent LCD module 30 insulation test voltages 500V 1kV 1 5kV 2kV 2 5kV 3kV 3 5kV 4kV 4 5kV 5kV 5 5kV 6kV 6 5kV 7kV 7 5kV 8kV 8 5kV 9kV 9 5kV 10kV 10 5kV...

Страница 6: ...t circuit current up to 5mA Current measurement 0 5nA 0 55mA Depending on the insulation resistance PI Polarization Index DAR Dielectric Absorption Ratio Voltmeter ACV 30 600V 50 60Hz DCV 30 600V Accu...

Страница 7: ...if needs cleaning DIRTY INSULATOR ELECTRICAL EQUIVALENT CIRCUIT DIRTY INSULATOR SECOND MEASUREMENT Measure with the guard to ensure insulator is correct TYPICAL TEST Resistance due to contamination ca...

Страница 8: ...V button 5 Insulation resistance test at 15kV button 6 To add 500V button to the selected test voltage 7 To subtract 500V button to the selected test voltage 8 TEST STOP button 9 ENTER SAVE button 10...

Страница 9: ...re that no voltage is present If voltage is present remove the power supplied 2 To secure operator s safety check if there is any damage on the tester or test cable 3 During the test do not touch any...

Страница 10: ...as below 5 Then press TEST STOP Note a During test process if there is a exterior voltage above AC30V or DC30V exerted beeper activates in response and LCD displays the warning picture as shown below...

Страница 11: ...EST button Do not press the TEST button in advance 7 1 3 Measure voltage Voltage Meter Function 1 1 Press ON OFF button Main page 2 Press value add or value reduce LCD displays the picture shown below...

Страница 12: ...STOP to switch to next time unit and go on the required adjustment 5 After all adjustments are complete press ENTER SAVE to confirm and save the data measured Note if time unit year month day hour min...

Страница 13: ...to confirm save the data measured 6 Press ESC to return back to the main page 7 1 6 Display the data stored LOG Display Function 4 1 Press ON OFF button Main page 2 Press value add 4 times LCD displa...

Страница 14: ...value add or value reduce to select the required data value If no data available LCD displays the picture shown below 9 Press ENTER SAVE to query the subpage of data 10 After the query is over press...

Страница 15: ...low Not erase press ESC to go back to the main page 7 1 8 Introduction of other Functions 1 Dielectric absorption ratio DAR Ratio of insulation resistance between 1 min and 30 sec 2 Polarization index...

Страница 16: ...eaches 16 5V charge process is complete It can be observed on LCD display No measurement can be performed during the charging process 9 Maintenance repair 9 1 To avoid and electric shock or device dam...

Страница 17: ...ation Steps with the USB Flash Drive 1 This program will install Insulation Tester on your computer automatically 2 Click the Next key to set 3 If you want to install a different folder click Browse a...

Страница 18: ...4 Click the Next key 5 It will show the information of all files are Installing to your personal computer 18...

Страница 19: ...icate the driver automatically It s necessary to install the driver if your computer system is not windows 7 then the driver is in the compact disk CD The directory is E USB DRIVER CDM 2 08 24 WHQL Ce...

Страница 20: ...3 Right click on Computer Click on Properties in the sub menu 4 The windows appears Control Panel Home On the left hand side of the window click on Device Manger 20...

Страница 21: ...the Run command window Start Menu Run or press R and type devmgmt msc without the quotation marks Click on the arrow to the left of Ports COM LPT to expand the listing 6 right click on the correct CO...

Страница 22: ...expand it and click on a COM port number within the range of COM1 to COM8 preferably one not in use avoid choosing COM1 if all are in use and there are no other peripherals connected to your PC assig...

Страница 23: ...manager 10 3 Insulation Tester Software Comm Port setting 1 Connect data transmission cable to the InsulationTester 2 Click the icon of the 15kV Insulation Tester on the desktop of the personal compu...

Страница 24: ...0 4 Insulation Tester Interface RS 232 Connection Main operation interface Memory saving and downloading interface Click the Download Log from Device key to download the current data and statistics 24...

Страница 25: ...Click the Chart Display key to see the chart picture shown in below 25...

Страница 26: ...Click the Save Log key to the file picture shown in below 26...

Страница 27: ...27...

Страница 28: ...www major tech com www majortech com au sales major tech com info majortech com au MAJOR TECH PTY LTD South Africa Australia...