5.6.17 Settings
– Service – General
Figure 134
Calibration reminder
: Annual calibration of the
touchTymp and its transducers is recommended.
Select or deselect this item to enable or disable a
reminder that will display daily. The reminder starts
1 month prior to the expiration of the calibration date for
your acoustic transducer(s) (Figure 134).
The user can always bypass the reminder message and
continue with screening.
Figure 135
Display calibration date
: Only for service (Figure 135).
Figure 136
Device calibration:
Only for service (Figure 136).
Figure 137
Restart to service mode:
Only for MAICO Technical
customer support (Figure 137).
Figure 138
Screen calibration:
Only for Service (Figure 138).
Figure 139
Test printout:
prints a test printout (without a session
Figure 140
Figure 141
Export error log:
If an error is occurring you can export
the error log data onto a USB flash drive (Figure 140). If
there is no USB flash drive connected the message box
(Figure 141) will be shown with further information.
NOTE:
Detection of the USB flash drive can take up to
10 seconds.
Figure 142
Figure 143
Application update:
for updating via USB flash drive
(Figure 142). A message box will be shown to ask you if
you want to continue updating (Figure 143).
NOTE:
The device will be restarted after application update.
Figure 144
Export Settings:
Export file onto USB flash drive (Figure 144).
Figure 145
Import Settings:
Import file from USB flash drive (Figure 145).
Figure 146
Reset Settings:
Resets settings to default. Facility
information will also be deleted (Figure 146).
Figure 147
VNC Server:
Only for service (Figure 147).