2. SAFETY PRECAUTIONS – IMPORTANT !
The following precautions should be always be exercised with the use of all electro-medical equipment to ensure
safety to all involved parties – user(s), patient(s), etc.
2.1 TRAINING
This equipment should only be used under the supervision of a trained physician in a medical facility. Do not use in
other locations or for any other purposes than the intended application.
2.2 INSTALLATION
1. This equipment should NEVER be installed or used in areas where the unit could get wet or be exposed to any
environmental conditions such as high temperature, humidity, direct sunlight, dust, salt, etc., which could
adversely affect the equipment.
2. This equipment should NEVER be installed or used in the presence of flammable or explosive gases or
chemicals.
3. This equipment should NEVER be installed, used or transported in an inclined position nor should it be
subjected to impact or vibration.
4. For safety reasons, this equipment must be properly grounded. (This equipment should be connected to a three
(3)-prong hospital grade receptacle in U.S.A. or Canada).
5. Ensure that all power requirements are met and comfort to those specified on the rating place located on the
rear panel.
6. Do not block the air intake vent of this equipment.
7. Do not allow the power cord to became twisted, crushed or pulled taut.
8. When using an isolation transformer for any ancillary equipment, ensure the power requirements of the devices
do not exceed the capacity of the isolation transformer. For further information, contact your local MAHE
distributor.
! WARNING
Never drop this equipment or subject it to severe impact as it could compromise the functionality and/or safety of
the unit. Should this equipment be mishandled or dropped, do not use it. Return it to an authorized MAHE
service facility for inspection and repair.
CAUTION:
All devices connecting to the MSV-2010 must be Classified Medical Equipment. Additional information
processing equipment connected to the MSV-2010 form a Medical System and the operator must determine that
all equipment comply with the appropriate end-product standard (such as IEC 60950 or IEC 60065) and the
Standard for Medical System, IEC 60601-1-1.
2.3 PRIOR TO USE
1. Confirm that this equipment functions properly and check the operation of all switches, indicators, etc.
2. To prevent electrical shock when used with endoscopes, this equipment is insulated (type BF electro-medical
equipment). Do not allow it to be grounded to other electrical devices being used on the patient. Rubber gloves
should always be worn to prevent grounding through user(s).
3. Confirm that other devices used in conjunction with equipment function properly and that these other devices
will not adversely affect the operation or safety of this equipment. If any component of the endoscopic system
is not properly functioning, the procedure should not be performed.
4. To prevent any potential electro-magnetic interference, do not use any kind of cellular phone near the camera.
5. Check and confirm that all cords or cables are connected correctly and securely.