5. INSTALLATION
5.1 SETTING UP THE MSV-2010
1. Place the video camera on a stable surface (cart, counter, stand, etc.).
NOTE
- Avoid places where the video camera may be splashed with liquid.
- Absolutely DO NOT use in any environment with explosive or flammable gases.
- DO NOT block the venting grids of the video camera
.
2. Make sure the power switch is in OFF position.
3. Plug the power cord into the power source using the three (3) prong plug supplied with the unit.
CAUTION - To prevent electric shock, connect power cords of peripheral equipment through
medical isolation transformers.
NOTE
- When using medical isolation transformer, be sure to check that the
transformer ratings.
Make sure that the power cord is connected to the mains with a three prong plug
(In USA use UL2601-1 rated isolation transformers and/or power strips only).
5.2 CONNECTING THE CAMERA HEAD
CAUTION
- When removing the camera head from the camera control unit, grasp the connector
area to release the lock mechanism. Pulling the cable could result in damages to the device.
NOTE
- Do not connect or disconnect the camera head during operation. It may cause
unrecoverable damage to the camera head.
To connect the camera head to the camera control unit align the red dot on the camera cable connector with the red
dot on the CAMERA jack, located on the left side of the camera control unit front panel. Gently push the connector
into the jack.
5.3 CONNECTING PERIPHERAL EQUIPMENT
1. The MSV-2010 can be connected to a video monitor via three types of video signals: composite video, Y/C
(S-VHS) or RGB.
It is possible to connect up to four appliances including the monitor simultaneously to the different video
outlets of the MSV-2010: VCRs, Image Processing PCs (via frame grabbers) and/or video printers.
2. Depending on the requirements of appliances you will be using, attach the appropriate cables to the rear panel
of the camera control unit. For example, if the monitor requires input from a Y/C cable, attach one end of the
Y/C cable to the corresponding connector on the rear panel of the camera control unit and the other end to the
monitor.
NOTE
- For highest quality pictures Y/C or RGB connections are recommended
CAUTION
- When used in clinical or residential areas near radio or TV units, this equipment
may be subjected to radio interference. To avoid adverse electromagnetic effects,
DO NOT operate this equipment near RF energy equipment.