10. TROUBLESHOOTING
Problem Solution
The power indicator
(refer to Figure 1) is not
lit.
A. Check that the AC power cable is properly connected.
. B. Check the camera control unit fuses. If necessary,
. replace with new fuses (type SB 2A).
The power indicator
(refer to Figure 1) is lit
but the monitor screen is
dark.
A. Check that all connections conform to the installation
instructions.
B. Make sure that the monitor and light source are turned
on.
C. Check the monitor contrast and brightness
controls, and make sure they are correctly set.
D. Dismount the camera from the endoscope, and
point the camera head towards a light source. If
the screen becomes brighter, check the
endoscopic equipment.
The picture is
monochromatic.
Check that your monitor is of the same standard
(PAL or NTSC) as the supplied video camera.
B. Check that that Chroma or Color control of your monitor
is not set to minimum.
C. Check the connection of the video cables on the rear
panel of the camera control unit (refer to Figure 2).
The picture is not in
focus.
A. Adjust the focus of the endoscope, if exists.
B. Focus the opto-mechanical adapter.
The picture is vertically
and/or horizontally
unstable.
A. Adjust the vertical and/or horizontal controls of the
.
monitor.