be significantly shorter if climatic conditions are unfavourable (i.e. high
humidity, extreme temperatures).
9.2
Cleaning the Sampling Probe and Tubing
Soot and condensate collects in the sampling tubing. These residues are removed
as follows:
1 Detach the sampling tubing from the unit.
2 Use compressed air (5 bar max.) to blow out the tubing towards the exhaust
probe until no more condensate and soot comes out.
9.3
Changing the Black F1 Filters
The MET-SERIES is supplied with the black F1 screw-in filters since the produc-
tion date 09/2019.
Maintenance kit contains 2x
1 Open the right maintenance flap
and remove the two F1 filters.