l  Stop sampling timer on analyser. 
m  The total sampling time is 60 seconds. The sampling must pick up rise and decay of the 

engine from idle to full throttle and return to idle. 

n  Remove the probe if no further testing is required. 
o  Record the mean (average) engine exhaust diesel particulate concentration in (mg/m

3

). 

 

 

Figure 1 – Diesel particulate load cycle – Standard method 

 

Alternative Method 

Where testing in accordance with clause 6.3.2.1 above is not practicable (such as engines fitted 
with manual clutches) diesel particulate testing of the undiluted raw exhaust of diesel engines 
shall be carried out as follows: 
a  Testing equipment shall comply with clause 8.2 

Diesel Particulate Monitoring Equipment. 

Note: 

It is not possible to use the Bosch Smoke Meter for the above tests due to its inability to 
measure over a 60 second sampling period. 
b  Exhaust sampling and analysis shall be carried out by a person competent to the satisfaction 

of the mine mechanical engineer to carry out such test or by a licensed laboratory. 

c  The diesel engine shall be operated until the engine temperature stabilises. Exhaust 

particulate sampling and analysis shall be performed when the diesel engine is at normal 
operating temperature. 

d  The vehicle must be safely chocked and the vehicle brakes must be applied. 

Important Safety Notice: 

As engines must be tested in the load condition requiring the drive train to be engaged during 
testing it is important the vehicle is adequately chocked or restrained and the brakes applied to 
ensure it cannot “jump” or run away while the diesel engine is under throttle during the test. 
e  Place vehicle in third gear or other gear as specified by the equipment manufacturer to carry 

out the test. 

f  Run the engine at idle speed. 
g  Insert sensor probe into the undiluted raw exhaust before any particulate treatment. 

Note:

 

This may be downstream of catalytic converters and water based scrubbers but must be before 
any particulate filter. 
h  Testing shall be carried out over a 60 second cycle and on the undiluted raw exhaust as 

indicated in  below. 

Notes: 

1  It is important that the timing of the idle and full throttle settings is as accurate as possible. 

To ensure good repeatability, it is advisable for the operator to practice the procedure 
several times prior to collecting a sample. 

2  Sampling should commence in the first idle period and continue until the final idle period. 

Содержание MPM-4M

Страница 1: ...rden soll Fehler Verwenden Sie die Registerkarte Start um Name dem Text zuzuweisen der hier angezeigt werden soll Fehler Verwenden Sie die Registerkarte Start um Name dem Text zuzuweisen der hier ange...

Страница 2: ...y 6 3 4 Sample In 7 3 5 Sample Out 7 3 6 RS 232 Output 7 3 7 Analog Output when installed 7 3 8 Remote Control Optional 8 4 Operation 9 4 1 Warm up 9 4 2 Auto Zero Adjust 9 4 3 Test Selection 10 4 4 S...

Страница 3: ...acturer In case of non compliance the declaration of conformity becomes void Any use other than described is inappropriate 1 4 Safety Instructions for Operation The accident prevention regulations of...

Страница 4: ...equire service at the manufacturer or distributor including possible replacement of expensive components The MPM 4M design ensures that both the laser and the detector are not only readily acces sible...

Страница 5: ...ates a number of features that enhance its usefulness in real world situations including Simple one button operation rugged construction for on site applications user maintainable with minimal tools v...

Страница 6: ...ond by second data capture and graphing software for PC 3 2 Instrument Setup Water Trap Sample Filter Fig 1 Rear Panel 3 3 Power Supply The Particle Measurement Unit requires a regulated 12 V DC power...

Страница 7: ...exits from this hose fitting 3 6 RS 232 Output The RS 232 interface can be used to integrate the analyzer into the optional MPM View software for second by second data logging and graphing on a PC see...

Страница 8: ...ch accounts for some minor deviations due to continuous minor fluctuations in PM sample concentration Fig 4 Volts vs PM Concentration mg m3 Low and High Ranges 3 8 Remote Control Optional If an option...

Страница 9: ...erating mode by simply pressing the red Enter push button switch on the front panel and holding it down for around two seconds During the warm up process the zero may be slowly drifting Ensure that th...

Страница 10: ...sting for longer periods under actual operating conditions hold down the ENTER button while turning on the power switch Real time DPM levels can be recorded via the MPM View software on a laptop This...

Страница 11: ...ENTER push button The green LED on the front panel will illuminate steady and a 5 second countdown will start allowing the operator to anticipate the time when measurements will commence Run Test Imm...

Страница 12: ...he supplied MPM View to generate a real time graph of the PM concentration profile over the complete test and record the second by second test data in an EXCEL compatible file See the supplied MPM Vie...

Страница 13: ...nt will start measuring and storing PM concentration measurements and the display will change to the following IDLE screen To assist the operator the green LED will flash slowly for 8 seconds indicati...

Страница 14: ...14 BA314001 en Full Power Idle Full Power Idle 12 seconds...

Страница 15: ...NOTE to log data in the continuous measuring mode the MPM 4M must be connected to a PC laptop with the MAHA Emission Viewer software running see separate user manual To initiate the continuous measuri...

Страница 16: ...ear panel of the analyser by rotating it counter clockwise and then pulling directly from its socket If the plug has not been removed for some time it may require a firm pull to overcome O ring fricti...

Страница 17: ...nd Correct Installation of Detector Head on the measuring chamber 7 After proceeding as above if the display reading is still not within the figure marked on the calibration plug unscrew and remove th...

Страница 18: ...e clear casing The filter should be checked daily Filtered sample flows from the inside to the outside of the media If black or dark brown discolouring is evident on the media surface the element must...

Страница 19: ...check is done monthly The following photograph shows the correct orientation for re fitting the disposable filter Fig 9 Internal Filter Orientation 7 Disposal If you want to dispose of the equipment p...

Страница 20: ...hicle must be safely chocked and the vehicle brakes must be applied Important Safety Notice As engines must be tested in the load condition requiring the drive train to be engaged during testing it is...

Страница 21: ...abilises Exhaust particulate sampling and analysis shall be performed when the diesel engine is at normal operating temperature d The vehicle must be safely chocked and the vehicle brakes must be appl...

Страница 22: ...timer indicates 32 seconds and quickly release throttle n Idle engine until timer indicates 40 seconds o Quickly apply full throttle again hold until timer indicates 48 seconds and quickly release th...

Страница 23: ...are However errors cannot be fully excluded Subject to technical change without notice Document Document BA314001 en Approval Date 27 10 2011 Manufacturer MAHA Maschinenbau Haldenwang GmbH Co KG Hoyen...

Отзывы: