6 Maintenance 

 

Apart from routine zeroing and occasional range calibration, the instrument requires very little 
maintenance. The only consumable items are the high-efficiency HEPA filters (internal and 
external) used, respectively, to clean the airflow protecting the optical system lenses and to 
remove particles from the sample stream before discharging to atmosphere. 
 

6.1 

Checking and Replacing the Rear Panel Filter 

A high performance HEPA filter, manufactured to remove >99.9% of all particles down to 
0.1micron size, is installed on the rear panel of the instrument. 
 

 

Fig. 8: Rear Panel Filter 

 
 
The filter cartridge should be examined for evidence of particle "breakthrough" on the surface of 
the filter media (the white cylinder inside the clear casing). 
The filter should be checked daily. Filtered sample flows from the inside to the outside of the 
media. If black or dark brown discolouring is evident on the media surface, the element must be 
replaced. The filter in the above picture has gone a long way past the point when it should have 
been replaced! 
To replace the media element, first unscrew and remove the clear plastic filter bowl. Then 
unscrew and remove the black plastic media retainer, together with the media element. Remove 
and replace the element with one of similar specification. 
Replacement and reassembly are the opposite sequence to above. When installing the clear 
plastic filter bowl, do not over-tighten the threads, as this may make subsequent removal 
difficult. Sealing is effected by the O-ring on the cylindrical body, not by end contact, so do not 
rotate the bowl further once resistance is felt. Occasionally lightly smear the O-ring with an 
appropriate lubricant. 
 

Содержание MPM-4M

Страница 1: ...rden soll Fehler Verwenden Sie die Registerkarte Start um Name dem Text zuzuweisen der hier angezeigt werden soll Fehler Verwenden Sie die Registerkarte Start um Name dem Text zuzuweisen der hier ange...

Страница 2: ...y 6 3 4 Sample In 7 3 5 Sample Out 7 3 6 RS 232 Output 7 3 7 Analog Output when installed 7 3 8 Remote Control Optional 8 4 Operation 9 4 1 Warm up 9 4 2 Auto Zero Adjust 9 4 3 Test Selection 10 4 4 S...

Страница 3: ...acturer In case of non compliance the declaration of conformity becomes void Any use other than described is inappropriate 1 4 Safety Instructions for Operation The accident prevention regulations of...

Страница 4: ...equire service at the manufacturer or distributor including possible replacement of expensive components The MPM 4M design ensures that both the laser and the detector are not only readily acces sible...

Страница 5: ...ates a number of features that enhance its usefulness in real world situations including Simple one button operation rugged construction for on site applications user maintainable with minimal tools v...

Страница 6: ...ond by second data capture and graphing software for PC 3 2 Instrument Setup Water Trap Sample Filter Fig 1 Rear Panel 3 3 Power Supply The Particle Measurement Unit requires a regulated 12 V DC power...

Страница 7: ...exits from this hose fitting 3 6 RS 232 Output The RS 232 interface can be used to integrate the analyzer into the optional MPM View software for second by second data logging and graphing on a PC see...

Страница 8: ...ch accounts for some minor deviations due to continuous minor fluctuations in PM sample concentration Fig 4 Volts vs PM Concentration mg m3 Low and High Ranges 3 8 Remote Control Optional If an option...

Страница 9: ...erating mode by simply pressing the red Enter push button switch on the front panel and holding it down for around two seconds During the warm up process the zero may be slowly drifting Ensure that th...

Страница 10: ...sting for longer periods under actual operating conditions hold down the ENTER button while turning on the power switch Real time DPM levels can be recorded via the MPM View software on a laptop This...

Страница 11: ...ENTER push button The green LED on the front panel will illuminate steady and a 5 second countdown will start allowing the operator to anticipate the time when measurements will commence Run Test Imm...

Страница 12: ...he supplied MPM View to generate a real time graph of the PM concentration profile over the complete test and record the second by second test data in an EXCEL compatible file See the supplied MPM Vie...

Страница 13: ...nt will start measuring and storing PM concentration measurements and the display will change to the following IDLE screen To assist the operator the green LED will flash slowly for 8 seconds indicati...

Страница 14: ...14 BA314001 en Full Power Idle Full Power Idle 12 seconds...

Страница 15: ...NOTE to log data in the continuous measuring mode the MPM 4M must be connected to a PC laptop with the MAHA Emission Viewer software running see separate user manual To initiate the continuous measuri...

Страница 16: ...ear panel of the analyser by rotating it counter clockwise and then pulling directly from its socket If the plug has not been removed for some time it may require a firm pull to overcome O ring fricti...

Страница 17: ...nd Correct Installation of Detector Head on the measuring chamber 7 After proceeding as above if the display reading is still not within the figure marked on the calibration plug unscrew and remove th...

Страница 18: ...e clear casing The filter should be checked daily Filtered sample flows from the inside to the outside of the media If black or dark brown discolouring is evident on the media surface the element must...

Страница 19: ...check is done monthly The following photograph shows the correct orientation for re fitting the disposable filter Fig 9 Internal Filter Orientation 7 Disposal If you want to dispose of the equipment p...

Страница 20: ...hicle must be safely chocked and the vehicle brakes must be applied Important Safety Notice As engines must be tested in the load condition requiring the drive train to be engaged during testing it is...

Страница 21: ...abilises Exhaust particulate sampling and analysis shall be performed when the diesel engine is at normal operating temperature d The vehicle must be safely chocked and the vehicle brakes must be appl...

Страница 22: ...timer indicates 32 seconds and quickly release throttle n Idle engine until timer indicates 40 seconds o Quickly apply full throttle again hold until timer indicates 48 seconds and quickly release th...

Страница 23: ...are However errors cannot be fully excluded Subject to technical change without notice Document Document BA314001 en Approval Date 27 10 2011 Manufacturer MAHA Maschinenbau Haldenwang GmbH Co KG Hoyen...

Отзывы: