2.2 PARASITIC FORCES
Incorrectly mounted torque transducers can generate parasitic forces on the measuring shaft in radial (F
R
)
and axial direction (F
A
).
Fig.2-4 Parasitic Forces
2.2.1 RADIAL FORCES (BENDING)
Radial forces (F
R
) generate a bending momentum in the measuring shaft resulting in displace-
ment of its center of gravity. This disequilibrium will load the shaft periodically with a frequency proportion-
al to the speed of rotation. This effect is particularly noticeable at high speeds.
IN EXTREME CASES, A HIGH BENDING FORCE MAY CAUSE PERMANENT DEFORMATION OF
THE MEASURING SHAFT, LEADING TO FALSE MEASURING RESULTS.
The following table lists the maximum radial forces F
R
allowed for TMB, TM and TMHS Series torque trans-
ducer shafts in suspended and supported installations.
MODEL
F
R
max.
(Suspended
installations)
F
R
max.
(Supported installations)
TM / TMB
a)
TMHS
N
N
N
TM 301
b)
8
N/A
TM 302
b)
16
TM 303
b)
25
25
TM 304
20
50
50
TM 305
40
80
80
TM 306
70
120
120
TM 307
60
120
TM 308
80
160
TM 309
60
150
150
TM 310
120
300
280
TM 311
200
410
TM 312
300
570
420
TM 313
500
550
410
TM 314
800
900
680
TM 315
1 100
850
640
TM 316
2 200
1 460
1 090
TM 317
1 300
980
d) If available (
see section 1.3 - Fiche technique
)
e) Suspended installation is not recommended for these models
F
R
F
A
F
A
F
A
F
R
F
R
F
R
Suspended installation
Supported installation
CAUTION
15
TM
series
INSTALLATION / CONFIGURATION