Magtrol Model 6510
e
Single-Phase Power Analyzer
Chapter 5 – Computer Controlled Operation
OPERA
TION
OPERA
TION
Command Code
function
explanation
MFm<terminator>
Sets the AC and DC
measurement mode filter.
"m" indicates the measurement filter low-
pass frequency (response time).
Values for
m
are:
0 = 1 Hz
1 = 2 Hz
2 = 5 Hz
3 = 10 Hz
4 = 20 Hz
5 = 40 Hz
6 = 100 Hz
PC<terminator>
Clears all peak hold values
(volt, amp and watt).
---
RA1,m2<terminator>
Sets current range and
input source.
"m2" indicates current range.
Values for
m2
are:
0 = 20 A
1 = 10 A
2 = 5 A
3 = 1 A
4 = 1 V external sensor input range
5 = 500 mV external sensor input range
6 = 250 mV external sensor input range
7 = 50 mV external sensor input range
RV1,m2<terminator>
Sets voltage range.
"m2" indicates voltage range.
Values for
m2
are:
0 = 600 V
1 = 300 V
2 = 150 V
3 = 30 V
SA1,m2<terminator>
Sets the amps scaling
constant (current
transformer).
"m2" indicates the amps scaling constant
in A/A and must be set within the following
range:
0.01 < m2 < 10000
When m2 = 0, the amps scaling mode will
be cleared.
SV1,m2<terminator>
Sets the voltage scaling
constant (potential
transformer).
"m2" indicates the voltage scaling
constant in V/V and must be set within the
following range:
0.01 < m2 < 10000
When m2 = 0, the voltage scaling mode
will be cleared.
SE1,m2<terminator>
Sets the external current
sensor scaling constant.
"m2" indicates the external current sensor
scaling constant in A/mV and must be set
within the following range:
0.0001 < m2 < 99999
note:
External sensor voltage input
values are divided by this value to give
output in amps in external sensor input
current ranges.