21
Magtrol Model 6510
e
Single-Phase Power Analyzer
Chapter 3 – Installation/Configuration
SETUP
SETUP
3.3.4
a
MP
S
caling
The current measurement range can be extended by using a current transformer. Frequency response
will be determined by the characteristics of the transformer used.
3.3..1
Hardware Connection
EARTH
LINE
HIGH
LOW
MOV 2
MOV 3
PT
CT
LOAD
CB
MOV 1
A
N
A
L
O
G
O
U
T
P
U
T
±
V
A
±
EXT.
SYNC.
EARTH
GROUND
20VA 50/60Hz
FUSE (5x20mm)
CAUTION: DOUBLE POLE FUSING
120V UL/CSA 250mA 250V SB
240V IEC 160mA 250V T
RS-232C
GPIB/IEEE-488
MAGTROL INC.
20 A
AMPS
750 V
VOLTS
MAX.
EXT.
MAX.
1000 V
MAX.
MAX.
±10 V
SENSOR
Figure 3–12 Current/Potential Transformer Connection
3.3.4.2
Software Configuration
To configure the 6510
e
amp scaling for a current transformer, complete the following steps utilizing
the USER MENU located on the front panel of the unit.
1. Turn on the 6510
e
. See
Section 3.1 – Powering Up the 6510e
.
2. Press ENTER button. The display should appear as shown in
Figure 2–3 Device Setup
Menu
.
3. Press
and
buttons until AMP SCALING appears in the display.
4. Press ENTER button. The display will appear as follows.
CURRENT
VOLTAGE
POWER/PF
Figure 3–13 Amp Scaling Setup Menu
5. To set amp scaling, press the RMS button under A1 and use the
pq
buttons in the
USER MENU to adjust the value.