background image

6.4 Start-Up for Gage-Mounted

LTM Transmitters

 

 

 

If the transmitter does not appear to calibrate properly, or has an erratic 

output,  check  the  deadband  of  the  sensor  tube  and  ensure  that  the 

float is within the active region of the probe.  The active region of the 

sensor  tube  is  typically  marked  with  20mA  and  4mA  stickers  when 

the unit is calibrated before shipment.

If  the  output  is  still  erratic,  try  disconnecting  the  power  momentarily 

by unplugging and re-plugging-in the terminals.  If a glitch was stored 

in RAM memory, this will generally clear it.  

It  is  possible  for  the  LTM-300  to  be  magnetically  biased  or  have 

residual  magnetic  energy  stored  along  the  length  of  the  waveguide.  

These magnetic anomalies can interfere with the signal-to-noise ratio 

and the stability of the output signal itself.

If  this  appears  to  be  the  case,  a  gage  float  magnet  (or  any  magnet 

available)  may  be  run  along  the  length  of  the  sensor  tube,  past  the 

head of the transmitter, in an even motion and without stopping.  This 

will usually clear all such magnetic anomalies.  

The LTM-300 is designed to operate with a supply voltage of 15-36 vdc 

(at 20 mA) across its terminals without affecting the mA signal. The most 

common loop supply used is 24vdc.

It is found sometimes that additional resistance in the loop is necessary, 

either in the form of a second load resistor or higher resistance safety 

barrier. This will appear to limit the maximum output of the transmitter 

to below 20 mA. The transmitter will generally perform correctly up to 

this point. To resolve this problem, the voltage of the loop supply must be 

increased somewhat. Even a slight increase by one volt may be sufficient 

and many supplies have such an adjustment.

All connections must be checked for improper wiring or polarity before 

power  is  applied. The  LTM-300  has  series  diodes  preventing  the 

reverse polarity from entering and damaging the circuitry. If power is 

applied  and  the  signal  is  0.0  mA,  chances  are  good  that  the  polarity 

is reversed.

 

SECTION 6.  Troubleshooting and Maintenance

6.0 Diagnostics are Via HART

6.1 Calibration Problems

6.2 Magnetic Interface

6.3 Troubleshooting Power Supply Problems

Never move a magnet in a perpendicular motion from along 

the sensor tube.  This will always leave a residual field in the 

waveguide which will cause the transmitter to give an erratic 

output

Gage and Transmitter Installation

1.  Visually inspect Mag-Gage and transmitter installation to 

insure transmitter is positioned with the 4 mA and 20 mA 
labels directly adjacent to the process connections.  Make sure 
transmitter is securely clamped and parallel to gage chamber.

2.  Remove bottom flange and install float.  Each float is clearly 

tagged with serial number of the Mag-Gage and process 
parameters.  Top of float is clearly marked “TOP” to insure float 
is in right side up.

3.  Inspect bottom flange for proper gasket and spring.  This 

spring protects float and keeps it from dropping below the 
bottom process connection.  To determine proper spring length 
measure “A” dimension of gage.  This is the length from the 
center of the bottom process connection to face of the bottom 
drain flange.  Length of spring should be

(“A” DIMENSION) (-) Minus (Float 2 inches)

       EXAMPLE:  “A” Dimension is 14”, Float length is 12”, + 2” = 4”
       Spring Length.

                                     14 – 12 + 2 = 4” spring

4.  Float is properly installed if the bottom 2-3 flippers on gage 

indicator have flipped.

 

Transmitter Check-out and Calibration

NOTE: The LTM Series Transmitters are 24VDC Loop powered (2-
wire) and require a minimum of 15 Volts at 20 mA. 

1.  Using a HART compatible loop calibrator, connect the “+” terminal 

on the LTM to positive lead of calibrator and the “-” terminal  on the 
LTM to negative lead of calibrator.

2.    With float in the gage at 4mA, the output of the LTM should be         
       4mA. Connect the HART Communicator to the transmitter. 
       Upon power up, the HART Communicator should read the LRV
       (lower range value) or 0 inches at 4mA.  

NOTE: There is a top spring in the Mag-Gage to 

protect float and prevent it from passing the top 

process connection.  DO NOT REMOVE EITHER 

SPRING.

3.  With float in the gage at 20mA, the output of the LTM should be 

20mA. HART should also display URV (upper range value) or 
span length in inches at 20mA.

4.  To insure complete functionality of gage and transmitter, fill 

the gage chamber with liquid and drain slowly to observe 
transmitter and gage are tracking properly.

6

CAUTION:

TERMINAL BOARD

Содержание LTM-300 Series

Страница 1: ...81 488 0788 FAX 281 488 7080 EMAIL magtech isemagtech com www isemagtech com LTM 300 Series Magnetostrictive Level Transmitter Instruction Operation Manual 3902 Magnolia Road Pearland Texas 77584 Jan...

Страница 2: ...r Installation 3 3 1 Standalone Transmitter Installation 3 3 2 Recommended Wiring Single Loop 4 Section 4 Speci cations Page 4 0 Transmitter Electrical Speci cations 4 4 1 Transmitter Sensor Tube 4 Se...

Страница 3: ...hange and continuously activates the sensor in the tube The electronics in the housing process the changing signal and update an analog 4 20 mA output This analog output is precisely proportional to t...

Страница 4: ...e base to resolve sub microsecond timing increments The binary number equivalent to the microseconds of the echo travel time is written to an output D A Converter and subsequently converted to a 4 20...

Страница 5: ...ectronicshousingmountedatthebottomofthegageforeasyaccess THIS MUST BE SPECIFIED AT TIME OF PURCHASE Installation Note European The cable entry devices and blanking elements of unused apertures shall b...

Страница 6: ...gurations A single output version with only one variable that will output a 4 20 mA signal for level The second con guration is a transmitter tted with a second oat to provide a digital output for in...

Страница 7: ...elect Data not required Span Trim Move the oat to the 20mA mark 100 on the probe Using the arrow buttons input the correct distance from the zero point and then Select Level O set Enter the desired va...

Страница 8: ...ion Problems 6 2 Magnetic Interface 6 3Troubleshooting Power Supply Problems Never move a magnet in a perpendicular motion from along the sensor tube This will always leave a residual eld in the waveg...

Страница 9: ...as this could leave magnetic indentation in the sensor wire The LTM 300 level transmitter has an inductor located inside the bulkhead of the sensor tube During operation of the transmitter this induc...

Страница 10: ...d the Alarm type In addition this screen presents the procedure to select the Primary variable PV if the transmitter is con gured with both Level and Interface It also displays the menu that will allo...

Страница 11: ...t are not normal Any status items not reported are within the normal range of operation or calibration The HART Communicator issues a command to receive the results of the transmitter diagnostics Duri...

Страница 12: ...xited if the user selects End The example shows the user selection 4mA The user must press ENTER before the procedure continues When the user selects the Loop test option the HART Communicator display...

Страница 13: ...output to 4 000mA the user is to enter the reference meter reading For example if the meter actually reads 4 234mA that value is input with the keypad The user must press ENTER when the current value...

Страница 14: ...matic control After the user presses OK the HART Communicator returns to the Calibration screen The HART Communicator is now at the starting screen for the Zero Level Trim Full Level Trim Starting at...

Страница 15: ...now be returned to automatic control After the user presses OK the HART Communicator returns to the Calibration screen The HART Communicator is now at the starting screen for the Full Level Trim Zero...

Страница 16: ...ed Once the measurement is stable the procedure will continue without user intervention The HART communicator will then send the command to the transmitter to calculate the new Temperature zero calibr...

Страница 17: ...t Value for 4mA and 20mA by Keypad If you desire to change the Level value at which the Current output is 4mA and 20mA when the Level is the selected Primary Variable or change the Interface value at...

Страница 18: ...g the edit box to continue with the reranging of the transmitter On returning to the Keypad rerange screen the values shown for LRV and URV are the changes that the user has made The user should check...

Страница 19: ...pper range value getting an update of the digital reading from the transmitter or leaving the 20mA range point as it was If the user accepts the measurement value as the 20mA loop current point the HA...

Страница 20: ...h is set in the factory when the unit is prepared for shipment The probe length is used as a reference for a number of diagnostics fuctions as well as setting the URL and LRL shown on the Keypad reran...

Отзывы: