6.2 Technical Specification

a. Main unit

Screen 

 

 

 

 

15” Colour LCD, 1024 x 768 pixels

Display

 

Bar graph range  

 

 

30µV – 30mV peak to peak

 

Bar graph resolution 

 

16 segments, logarithmic 

 

Waveform amplitude ranges 

±25µV, ±50µV, 

±100μV (default)

, ±200µV, ±500µV,  

 

 

 

 

 

 

±1mV,  ±2.0mV, ±5mV, ±10mV, ±20mV, ±30mV 

 

Accuracy 

  

 

 

2% of FSD 

 

Waveform time base ranges  

12.5ms, 

25ms (default)

, 50ms, 

 

     100ms, 

500ms, 

1s, 

5s, 

10s 

 

Accuracy 

  

 

 

2% 12.5ms-1s; 5% 5s; 10% 10s. 

 

Time base cursor accuracy  

2% 

Software 

    See 

"6.4. System Details" on page 47 

Rotary 

actuators 

   3:

 

Left dial  

- Primary stimulator stimulation setting (STIM1)

 

Right dial  

- Secondary stimulator stimulation setting (STIM2)

 

Centre dial   - EMG response audio volume (Master Volume)

External connections 

 

 

EMG Amplifier Module (Pre-amplifier)

 

     Trigger 

In 

  

 

     Trigger 

Out 

– 

Open 

Drain 

   

 

     Mute 

sensor 

   

 

     USB 

 

   

EMG Audio  

 

 

 

8.4W rms (95dBa @ 1m)

b. Stimulator pod

 Primary Stimulator

 

Operational mode   

 

Constant Current or Constant Voltage 

 

Current ranges 

 

 

10µA – 10mA, in steps dependent on procedure.

 

Accuracy  

 

 

 

±5%, into 1kΩ load

Содержание NEUROSIGN V4

Страница 1: ...THE NEUROSIGN V4 INTRAOPERATIVE NERVE MONITOR PRESERVING NERVE INTEGRITY Service Manual NOP20 EN 03 January 2021 2797...

Страница 2: ...n Company details Neurosign The Magstim Company Limited Spring Gardens Whitland Carmarthenshire SA34 0HR Website http www neurosignsurgical com Product Enquiries Telephone 44 0 1994 240798 Fax 44 0 19...

Страница 3: ...icing of the products named in this manual or otherwise not carried out by The Magstim Company Limited or persons authorised by the company will invalidate the product guarantee Within the guarantee p...

Страница 4: ...The Magstim Company Ltd iv NOP20 EN 03 Page intentionally left blank...

Страница 5: ...2 Disassembly 26 CARE MAINTENANCE 29 4 1 Cleaning Disinfecting 30 4 2 Storage Environment 30 4 3 User Maintenance Calibration 30 4 4 Servicing Device Lifetime 31 4 5 Packaging Transportation 32 4 6 D...

Страница 6: ...The Magstim Company Ltd vi NOP20 EN 03 Page intentionally left blank...

Страница 7: ...Neurosign V4 Pre amplifier 4 Channel P N 4444 00 Neurosign V4 Pre amplifier 8 Channel P N 4448 00 Neurosign V4 Mute Sensor P N 4225 00 Note Full information regarding consumables such as stimulating...

Страница 8: ...The Magstim Company Ltd 8 NOP20 EN 03 Page intentionally left blank...

Страница 9: ...must be filed promptly with the transportation company All correspondence concerning the equipment must specify the model name and or number as well as the serial number exactly as they appear on the...

Страница 10: ...The Magstim Company Ltd 10 NOP20 EN 03 Page intentionally left blank...

Страница 11: ...The Magstim Company Ltd 11 NOP20 EN 03 SECTION 1 PRODUCT OVERVIEW...

Страница 12: ...hold for 2 seconds to mute the EMG audio 4 USB Export Import The USB memory stick connection for import export provided with a rubber cover 5 Stimulator STIM 2 Dial Not currently available For more i...

Страница 13: ...Connection Connection socket for the pre amplifier lead 13 Slave Screen Connection For more information contact the product enquiries team page ii 14 Display Connection a Touch Screen DCU b Mains unit...

Страница 14: ...channel four orange 23 Reference Electrode Channel The corresponding sockets for the reference electrode channel green 24 Electrode Channels 5 8 The corresponding sockets for channel five grey channel...

Страница 15: ...of cable attached to the stimulating probes making them easier to handle 27 Mounting Clip A mounting clip is located on the side of the stimulator pod so that it can be attached to the side rail of th...

Страница 16: ...The Magstim Company Ltd 16 NOP20 EN 03 Page intentionally left blank...

Страница 17: ...The Magstim Company Ltd 17 NOP20 EN 03 SECTION 2 WARNINGS SYMBOLS...

Страница 18: ...equipment and result in improper operation WARNING Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches...

Страница 19: ...as this may cause damage In the event of hardware failure or software corruption the nerve monitor is designed to fail in a safe state In this state the stimulator and EMG audio channels are disabled...

Страница 20: ...ossible from the electrode leads Patients with active medical devices such as pacemakers or cochlear implants should not be stimulated using the stimulating probe without due consideration being given...

Страница 21: ...ements of the applicable CE mark EC directives Manufacturer see page ii for contact details DO NOT dispose in general waste Please contact the Magstim company Ltd for advise on disposal in compliance...

Страница 22: ...Magstim Company Ltd 22 NOP20 EN 03 Trigger input socket see page 13 Mute sensor socket see page 13 Ethernet socket see page 13 Stimulator pod to pre amplifier connection see page 15 USB socket see pag...

Страница 23: ...The Magstim Company Ltd 23 NOP20 EN 03 SECTION 3 ASSEMBLY DISASSEMBLY...

Страница 24: ...The Magstim Company Ltd 24 NOP20 EN 03 3 1 Assembly 6 6 5 5 4 4 60mm x 4 3 3 8mm x 4 2 2 70mm x 4 1 1 70mm x 4...

Страница 25: ...The Magstim Company Ltd 25 NOP20 EN 03 12 12 11 11 10 10 9 9 8 8 7 7...

Страница 26: ...ut socket right Remove the two bolts on either side of the mains unit below using a 5 mm allen key supplied with the nerve monitor Slide the rubber infill upwards Lift the mains unit upwards The mains...

Страница 27: ...data Do not send the SD card with the equipment i SD card removal Remove the retaining screw from the rear panel of the touch screen DCU right Remove rear panel by sliding downwards indicated by an a...

Страница 28: ...ard when the SD card is in place Note The SD card can only be inserted in one orientation Do not apply force or damage may occur Replace the rear panel and secure with the retaining screw Connect the...

Страница 29: ...The Magstim Company Ltd 29 NOP20 EN 03 SECTION 4 CARE MAINTENANCE...

Страница 30: ...ith hospital protocol The Magstim company Ltd recommends that a lint free cloth moistened with 70 isopropyl alcohol should be used Care must be taken to ensure the equipment is not soaked and has drie...

Страница 31: ...guarantee Performance verification should be carried out by service personnel annually once the guarantee period has expired Authorised service personnel should refer to Servicing on page 33 b Damage...

Страница 32: ...e last date of sale Spare parts cannot be guaranteed after this time Please contact the Magstim Company Ltd for information on spare parts availability after the expiration of the device lifetime The...

Страница 33: ...The Magstim Company Ltd 33 NOP20 EN 03 SECTION 5 SERVICING...

Страница 34: ...nel by sliding downwards indicated by an arrow Remove the button cell battery from the retaining clip right and dispose of appropriately Insert the new button cell battery ensuring the positive is fac...

Страница 35: ...ower switch to ON I 20 page 13 Note If language selection flags are displayed on start up the system configuration must be completed Refer to the user manual 4 1 System Configuration for more informat...

Страница 36: ...d and retry the update Message box Application is corrupted is displayed during start up below Remove the SD card from the nerve monitor and using a suitable SD card reader connect the SD card to a PC...

Страница 37: ...ate screen will now be displayed A progress bar and a label containing the message retrieving page sizes will be displayed The displayed message will then change to programming flash Oncompletionamess...

Страница 38: ...ld be named 4247 62 Vn n DCU bin where n n represents the software version of the update being applied If the correct file was chosen and the problem persists remove the USB memory stick from nerve mo...

Страница 39: ...ode 1478963 and press The system update screen will now be displayed Insert the USB stick into the external USB socket right Select 4247 62 Vn n BIN Where n n represents the software version Press nex...

Страница 40: ...hould be named 4248 62 Vn n bin where n n represents the software version of the update being applied If the correct file was chosen and the problem persists remove the USB memory stick from nerve mon...

Страница 41: ...vided at the bottom of the monitor case right Replace rear panel and secure with the retaining screw Set the power switch to ON I 20 page 13 To set the Keyboard language 5 4 Keyboard Installation Reta...

Страница 42: ...Ltd 42 NOP20 EN 03 If device schematics or drawings are required please contact the service department Servicing Enquiries Telephone 44 0 1994 242900 E mail service neurosignsurgical com 5 5 Device S...

Страница 43: ...The Magstim Company Ltd 43 NOP20 EN 03 SECTION 6 TECHNICAL INFORMATION...

Страница 44: ...re information Note To avoid problems with EMC interference the nerve monitor should not be used in the vicinity of any equipment that does not comply with EMC Safety Standard EN 60601 1 2 including m...

Страница 45: ...Accuracy 2 12 5ms 1s 5 5s 10 10s Time base cursor accuracy 2 Software See 6 4 System Details on page 47 Rotary actuators 3 Left dial Primary stimulator stimulation setting STIM1 Right dial Secondary s...

Страница 46: ...nels 4 Signal input range 5 V to 40mV Common mode rejection ratio 90dB 50 60Hz Bandwidth 10Hz 1kHz 3dB points Noise 20 V rms Input referred Notch filter 50Hz 60Hz off 6 3 General Specification Power 1...

Страница 47: ...ails screen 1 Serial Number The product serial number 2 Reference Number The product reference number 3 Date Of Manufacture The date this product was manufactured 4 Base Unit The details for the produ...

Страница 48: ...The Magstim Company Ltd 48 NOP20 EN 03 Page intentionally left blank...

Страница 49: ...Document Description Language NOP20 EN 02 Neurosign V4 Service Manual English UK Supporting documents Document Description Language NOP08 EN Neurosign V4 User Manual English UK NOP18 EN Neurosign V4...

Страница 50: ...www magstim com The Magstim Company Ltd...

Отзывы: