The Magstim Company Ltd.©
34
NOP20-EN-03
On start-up the nerve monitor will automatically run a connection and audio test.
Refer to the user manual for more information.
The nerve monitor must undergo annual electrical safety testing in accordance with
the hospital policy, as well as annual performance verification, once the gurantee
period has expired. More information can be found in "4.4 Servicing & Device Lifetime"
on page 31.
5.1 Testing Requirements
• Remove the retaining screw from the rear
panel of the touch screen DCU (
right
).
• Remove rear panel by sliding downwards
(indicated by an arrow).
• Remove the button cell battery from
the retaining clip (
right
) and dispose of
appropriately.
• Insert the new button cell battery ensuring
the positive (+) is facing outward.
• Replace the rear panel and secure with
retaining screw.
5.2 Battery Replacement
Retaining
Screw
Button Cell
Battery
• Connect the AC mains power lead to the power input socket (
19
• Set the power switch to ON (I) (
20
Note
: If language selection flags are displayed on start-up, the system configuration
must be completed. Refer to the user manual, "4.1 System Configuration" for more
information.
Note:
Before removing the retaining screws, ensure that the nerve monitor is switched
off and disconnected from the AC mains supply.
Содержание NEUROSIGN V4
Страница 4: ...The Magstim Company Ltd iv NOP20 EN 03 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP20 EN 03 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd 8 NOP20 EN 03 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP20 EN 03 Page intentionally left blank...
Страница 11: ...The Magstim Company Ltd 11 NOP20 EN 03 SECTION 1 PRODUCT OVERVIEW...
Страница 16: ...The Magstim Company Ltd 16 NOP20 EN 03 Page intentionally left blank...
Страница 17: ...The Magstim Company Ltd 17 NOP20 EN 03 SECTION 2 WARNINGS SYMBOLS...
Страница 23: ...The Magstim Company Ltd 23 NOP20 EN 03 SECTION 3 ASSEMBLY DISASSEMBLY...
Страница 24: ...The Magstim Company Ltd 24 NOP20 EN 03 3 1 Assembly 6 6 5 5 4 4 60mm x 4 3 3 8mm x 4 2 2 70mm x 4 1 1 70mm x 4...
Страница 25: ...The Magstim Company Ltd 25 NOP20 EN 03 12 12 11 11 10 10 9 9 8 8 7 7...
Страница 29: ...The Magstim Company Ltd 29 NOP20 EN 03 SECTION 4 CARE MAINTENANCE...
Страница 33: ...The Magstim Company Ltd 33 NOP20 EN 03 SECTION 5 SERVICING...
Страница 43: ...The Magstim Company Ltd 43 NOP20 EN 03 SECTION 6 TECHNICAL INFORMATION...
Страница 48: ...The Magstim Company Ltd 48 NOP20 EN 03 Page intentionally left blank...
Страница 50: ...www magstim com The Magstim Company Ltd...