background image

9

A G E N C Y   A P P R O V A L S

These devices are in compliance with the RED-directive 2014/53/EU, the EMC directive 2014/30/EU,
the PED-directive 2014/68/EU, the ATEX directive 2014/34/EU and RoHS directive 2011/65/EU.

Explosion Proof

US/Canada:

FM17US0108X / FM17CA0055X
Class I, Div 1, Group B, C, D, T4…T1
Zone 1 A Ex db ia IIB+H2 T4…T1
Zone 1 Ex d ia IIB+H2 T4…T1
Ta = -40 ºC to +70 ºC
Type 4X, IP67

Flame Proof 

ATEX – FM17ATEX0027X

II 1/2 G Ex db ia IIB + H2 T4…T1 Ga/Gb
Ta = -40 ºC to +70 ºC
IP67

IEC- IECEx FMG 17.0012X 

Ex db ia IIB + H2 T4…T1 Ga/Gb
Ta = -40 ºC to +70 ºC
IP67

Non- Incendive

US/Canada:

FM17US0108X / FM17CA0055X
Class I, II, III, Div 2, Group A, B, C, D, E, F, G, T4…T1 
Class 1, Zone 2 AEx nA ia IIC T4…T1    
Class 1, Zone 2 Ex nA ia IIC T4…T1    
Ta = -15 ºC to +70 ºC
Type 4X, IP67

ATEX - FM17ATEX0028X

II  3 G Ex nA  IIC Gc T4…T1 
Ta = -15 ºC to +70 ºC
IP67

IEC – IECEx FMG 17.0012X

Ex nA IIC Gc T4…T1
Ta = -15 ºC to + 70 ºC
IP67

Intrinsically Safe

US/Canada:

FM17US0108X / FM17CA0055X
Class I, II, III, Div 1, Group A, B, C, D, E, F, G, T4…T1
Class I, Zone 0 AEx ia IIC T4…T1
Class I, Zone 0 Ex ia IIC T4…T1 Ga 
Ta =-40 ºC to +70 ºC
Type 4X, IP67

ATEX – FM17ATEX0027X:

II 1 G Ex ia IIC T4…T1 Ga 
Ta = -40 ºC to +70 ºC
IP67

IEC – IECEx FMG 17.0012X:

Ex ia IIC T4…T1 Ga 
Ta = -40 ºC to +70 ºC
IP67

Dust Ignition Proof 

US/Canada:

FM17US0108X / FM17CA0055X
Class II, III, Div 1, Group E, F, and G, T4…T1 
Ta = -15º C to +70º C
Type 4X, IP67

ATEX – FM17ATEX0027X:

II 2 D Ex ia tb IIIC T100º C Db
Ta = -15º C to +70º C
IP67

IEC – IECEx FMG 17.0012X:

Ex ia tb IIIC T100º C Db
Ta = -15º C to +70º C
IP67

FM3600:2011, FM3610:2010, FM3611:2004, FM3615:2006, FM3616:2011, FM3810:2005, ANSI/ISA60079-0:2013,
ANSI/ISA 60079-1:2015, ANSI/ISA 60079-11:2013, ANSI/ISA 60079-15:2012, ANSI/ISA 60079-26:2011, NEMA 250:2003,
ANSI/IEC 60529:2004, C22.2 No. 0.4:2009, C22.2 No. 0.5:2008, C22.2 No. 30:2007, C22.2 No. 94:2001, C22.2 No. 213:2012,
C22.2 No. 1010.1:2009, CAN/CSA 60079-0:2011, CAN/CSA 60079-1:2011, CAN/CSA 60079-11:2014, CAN/CSA 60079-
15:2012, C22.2 No. 60529:2005, EN60079-0:2012, EN60079-1:2014, EN60079-11:2012, EN60079-15:2010, EN60079-26:2007,
EN60079-31:2009, A1:1991-2000, IEC60079-0:2011, IEC60079-1:2014, IEC60079-11:2011, IEC60079-15:2010,
IEC60079-26:2006, IEC60079-31:2008

“This equipment with chargeable non-conductive parts, e.g. enclosure’s paint and antenna use PTFE, Co-polymer
Polypropylene or Noryl En265, is provided with a warning label referring to the safety measures that must be taken if there is
electrostatic charging during operation. For use in hazardous area, the equipment and side to be installed, e.g. tank, must be
connected to earth and be attention to not only the measuring object, e.g. liquids, gases, powders and etc., but also the related
conditions, e.g. tank container, vessel and etc. (According to IEC 60079- 32-1).”

FCC (ID# LPN-R86) Compliance Statement:

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This
equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruc-
tions, may cause harmful interference to radio communications.

Telecommunications Approvals

Agency

In-Tank

Out of Tank

FCC

47 CFR, Part 15, Subpart C, Section 15.209

Unintentional Radiators

47 CFR, Part 15, Subpart C, Section 15.256

ISED

RSS-211

RSS-211

ETSI

EN 302 372 V2.1.1 (2016-12)

EN 302 729 V2.1.1 (2016-12)

Содержание Pulsar R86

Страница 1: ...s configuration wizards bring simplicity to an often complex technology This latest entry into the radar level measurement field is designed to provide unparalleled performance and ease of use The 26...

Страница 2: ...ght multiplied by the transit time divided by two Distance C Transit Time 2 Level is then calculated by factoring in tank height and other configuration information Figure 3 The reference point for di...

Страница 3: ...Turbulence Medium or Heavy Dielectric 1 7 3 3 10 10 100 1 7 3 3 10 10 100 Antenna Type 11 2 Horn 30 9 40 12 60 18 10 3 16 5 26 8 2 Horn 33 10 49 15 66 20 10 3 20 6 33 10 3 Horn 50 15 66 20 98 30 13 4...

Страница 4: ...maximum liquid surface with minimum striking of other objects in the vessel including the tank wall Use these drawings to determine the optimum installation location O B S T R U C T I O N S Almost any...

Страница 5: ...perature exposure to the transmitter in high temperature applications a heat extension 032 6922 001 is required between the antenna and transmitter Refer to Operating Temperature Range chart on page 6...

Страница 6: ...PTFE POLYPROPYLENE QUARTZ Ambient Temperature F C 40 40 20 29 0 18 20 7 40 4 60 16 80 27 100 38 120 49 140 60 160 71 180 82 700 371 750 400 600 315 500 260 400 204 300 149 200 93 100 38 0 18 100 73 P...

Страница 7: ...cannot be handled by the R86 transmitter Unusual flow applications can readily be accomplished through the use of either the 20 point Custom Table or the Generic Discharge Flow Equation which allows a...

Страница 8: ...C General purpose ethylene Ketones MEK acetone skydrol fluids amines anhy drous ammonia low molecular weight esters and ethers hot hydrofluoric or chlorosulfuric acids sour HCs Kalrez 4079 2 400 F 23...

Страница 9: ...1 2013 ANSI ISA 60079 15 2012 ANSI ISA 60079 26 2011 NEMA 250 2003 ANSI IEC 60529 2004 C22 2 No 0 4 2009 C22 2 No 0 5 2008 C22 2 No 30 2007 C22 2 No 94 2001 C22 2 No 213 2012 C22 2 No 1010 1 2009 CAN...

Страница 10: ...HART DD English French German Spanish Russian Chinese Portuguese FOUNDATION fieldbus Host System English Voltage Measured at instrument terminals HART General Purpose Weather proof Intrinsically Safe...

Страница 11: ...45 over entire temperature range 40 to 175 F 40 to 80 C Maximum Rate of Change 180 inches 450 cm minute FOUNDATION fieldbus ITK Version 6 2 0 H1 Device Class Link Master LAS selectable ON OFF H1 Profi...

Страница 12: ...I F I C A T I O N S I N C H E S M M 12 TRANSMITTER Model Number 3rd Digit Horn Size 11th Digit Extension 1 11 2 2 2 3 3 4 4 Dim H 0 None 3 2 81 4 5 114 8 5 216 11 5 292 Dim L 1 4 6 152 2 8 8 203 8 3...

Страница 13: ...ne FOUNDATION fieldbus only 5th digit 2 1 SIL 2 Hardware HART only 5th digit 1 A None FOUNDATION fieldbus only 5th digit 2 ETSI B SIL 2 Hardware HART only 5th digit 1 ETSI 7 ACCESSORIES MOUNTING 0 No...

Страница 14: ...150 ASME raised face flange EA DN 80 PN 16 EN 1092 1 Type A 54 3 300 ASME raised face flange EB DN 80 PN 25 40 EN 1092 1 Type A 55 3 600 ASME raised face flange ED DN 80 PN 63 EN 1092 1 Type B2 63 4...

Страница 15: ...NS 0 Viton GFLT 2 Kalrez 4079 8 Simriz SZ485 formerly Aegis PF128 NACE A Kalrez 6375 N None Quartz seal mandatory for flame proof ATEX IEC approval 0 None 1 For nozzle height 4 100 mm available only w...

Страница 16: ...o be cov ered under the warranty then MAGNETROL will repair or replace the control at no cost to the purchaser or owner other than transportation MAGNETROL shall not be liable for misap plication labo...

Отзывы: