background image

P R O B L E M A T I C   A P P L I C A T I O N S

G U I D E D   W A V E   R A D A R   A L T E R N A T I V E

Some applications can be problematic for Non-Contact

Radar. The following are examples of when Guided

Wave Radar is recommended.

Extremely low dielectric media (

ε

r<1.7)

Very weak reflections from the liquid surface 

(particularly during turbulence) can cause poor 

performance.

Tanks heavily cluttered with false targets (mixers,

pumps, ladders, pipes, etc.)

During times of very low liquid levels of low dielec-

tric media, the metal tank bottom may be detected,

which can deteriorate performance.

Foam can either absorb or reflect the microwave

energy depending upon the depth, dielectric, density

and wall thickness of the bubbles. Due to typical

variations in the amount (depth) of foam, it is

impossible to quantify performance. It may be

possible to receive most, some or none of the

transmitted energy.

Extremely high liquid level (Overflow) conditions

when liquid very near the antenna can cause

erroneous readings and measurement failure.

Interface applications

Refer to ECLIPSE Model 706 Guided Wave Radar 

bulletin 57-106.

8

Material Code

Maximum

Temperature

Maximum

Pressure

Min.

Temp.

Recommended

For Use In

Not Recommended

For Use In

Viton

®

GFLT

0

+400 °F @ 232 psi

(+200 °C @ 16 bar)

750 psi @ +70 °F

(51.7 bar @ +20 °C)

-40 °F

(-40 °C)

General purpose, ethylene

Ketones (MEK, acetone), 
skydrol fluids, amines, anhy-
drous ammonia, low molecular
weight esters and ethers, hot
hydrofluoric or chlorosulfuric
acids, sour HCs

Kalrez

®

4079

2

+400 °F @ 232 psi

+(200 °C @ 16 bar)

750 psi @ +70 °F

(51.7 bar @ +20 °C)

-40 °F

(-40 °C)

Inorganic and organic acids 
(including HF and nitric), 
aldehydes, ethylene, glycols,
organic oils, silicone oils, vinegar,
sour HCs

Black liquor, hot water/steam,
hot aliphatic amines, ethylene
oxide,propylene oxide, molten
sodium, molten potassium

Simriz SZ485

(formerly

Aegis PF128)

8

+400 °F @ 232 psi

+(200 °C @ 16 bar)

750 psi @ +70 °F

(51.7 bar @ +20 °C)

-4 °F

(-20 °C)

Inorganic and organic acids
(including HF and nitric), 
aldehydes, ethylene, glycols,
organic oils, silicone oils, vinegar,
sour HCs, steam, amines, ethyl-
ene oxide, propylene oxide,
NACE applications

Black liquor, Freon 43, Freon 75,
Galden, KEL-F liquid, molten
sodium, molten potassium

Kalrez

®

6375

A

400 °F @ 232 psi

(200 °C @ 16 bar)

750 psi 70 °F

(51.7 bar @ 20 °C)

-40 °F

(-40 °C)

Inorganic and organic acids
(including hydro fluids and nitric),
aldehydes, ethylene, organic oils,
glycols, silicone oils, vinegar,
sour HCs

Hot water/steam, hot aliphatic
amines, ethylene oxide, propylene
oxide

Quartz

N

+750 °F @ 1375 psi

(+400 °C @ 94.8 bar)

2320 psi @ 70 °F

(160 bar @ +20 °C)

-100 °F

(-70 °C)

General high temperature/high
pressure applications,
hydrocarbons, full vacuum
(hermetic), ammonia, chlorine

Hot alkaline solutions HF acid,
media with ph>12, direct exposure
to saturated steam

M300 °F (+150 °C) for use on steam.

O - R I N G   ( S E A L )   S E L E C T I O N   C H A R T

Содержание Pulsar R86

Страница 1: ...s configuration wizards bring simplicity to an often complex technology This latest entry into the radar level measurement field is designed to provide unparalleled performance and ease of use The 26...

Страница 2: ...ght multiplied by the transit time divided by two Distance C Transit Time 2 Level is then calculated by factoring in tank height and other configuration information Figure 3 The reference point for di...

Страница 3: ...Turbulence Medium or Heavy Dielectric 1 7 3 3 10 10 100 1 7 3 3 10 10 100 Antenna Type 11 2 Horn 30 9 40 12 60 18 10 3 16 5 26 8 2 Horn 33 10 49 15 66 20 10 3 20 6 33 10 3 Horn 50 15 66 20 98 30 13 4...

Страница 4: ...maximum liquid surface with minimum striking of other objects in the vessel including the tank wall Use these drawings to determine the optimum installation location O B S T R U C T I O N S Almost any...

Страница 5: ...perature exposure to the transmitter in high temperature applications a heat extension 032 6922 001 is required between the antenna and transmitter Refer to Operating Temperature Range chart on page 6...

Страница 6: ...PTFE POLYPROPYLENE QUARTZ Ambient Temperature F C 40 40 20 29 0 18 20 7 40 4 60 16 80 27 100 38 120 49 140 60 160 71 180 82 700 371 750 400 600 315 500 260 400 204 300 149 200 93 100 38 0 18 100 73 P...

Страница 7: ...cannot be handled by the R86 transmitter Unusual flow applications can readily be accomplished through the use of either the 20 point Custom Table or the Generic Discharge Flow Equation which allows a...

Страница 8: ...C General purpose ethylene Ketones MEK acetone skydrol fluids amines anhy drous ammonia low molecular weight esters and ethers hot hydrofluoric or chlorosulfuric acids sour HCs Kalrez 4079 2 400 F 23...

Страница 9: ...1 2013 ANSI ISA 60079 15 2012 ANSI ISA 60079 26 2011 NEMA 250 2003 ANSI IEC 60529 2004 C22 2 No 0 4 2009 C22 2 No 0 5 2008 C22 2 No 30 2007 C22 2 No 94 2001 C22 2 No 213 2012 C22 2 No 1010 1 2009 CAN...

Страница 10: ...HART DD English French German Spanish Russian Chinese Portuguese FOUNDATION fieldbus Host System English Voltage Measured at instrument terminals HART General Purpose Weather proof Intrinsically Safe...

Страница 11: ...45 over entire temperature range 40 to 175 F 40 to 80 C Maximum Rate of Change 180 inches 450 cm minute FOUNDATION fieldbus ITK Version 6 2 0 H1 Device Class Link Master LAS selectable ON OFF H1 Profi...

Страница 12: ...I F I C A T I O N S I N C H E S M M 12 TRANSMITTER Model Number 3rd Digit Horn Size 11th Digit Extension 1 11 2 2 2 3 3 4 4 Dim H 0 None 3 2 81 4 5 114 8 5 216 11 5 292 Dim L 1 4 6 152 2 8 8 203 8 3...

Страница 13: ...ne FOUNDATION fieldbus only 5th digit 2 1 SIL 2 Hardware HART only 5th digit 1 A None FOUNDATION fieldbus only 5th digit 2 ETSI B SIL 2 Hardware HART only 5th digit 1 ETSI 7 ACCESSORIES MOUNTING 0 No...

Страница 14: ...150 ASME raised face flange EA DN 80 PN 16 EN 1092 1 Type A 54 3 300 ASME raised face flange EB DN 80 PN 25 40 EN 1092 1 Type A 55 3 600 ASME raised face flange ED DN 80 PN 63 EN 1092 1 Type B2 63 4...

Страница 15: ...NS 0 Viton GFLT 2 Kalrez 4079 8 Simriz SZ485 formerly Aegis PF128 NACE A Kalrez 6375 N None Quartz seal mandatory for flame proof ATEX IEC approval 0 None 1 For nozzle height 4 100 mm available only w...

Страница 16: ...o be cov ered under the warranty then MAGNETROL will repair or replace the control at no cost to the purchaser or owner other than transportation MAGNETROL shall not be liable for misap plication labo...

Отзывы: