6
58-603 Pulsar
®
Model R86 Radar Transmitter
1.1.2 Configuration Information
A helpful SETUP WIZARD, which will guide you through
the simple configuration (with parameter explanations), is
available in the PULSAR Model R86. Located in the local
user interface menu under MAIN MENU/WIZARD/
SETUP WIZARD, some key information is required for
configuration. The transmitter will prompt confirmation
questions at the end of the Setup Wizard to verify operation.
Gather the information and complete the following operat-
ing parameters table before beginning configuration.
NOTE: These configuration steps are not necessary if the transmitter
was pre-configured prior to shipment.
Display
Question
Answer
Measurement What is the intended measurement
Type
type (Level, Volume, or Flow)?
_____________
System
What units of measurement will be
Units
used?
_____________
Antenna
What type of antenna is being used?
Model
Select first 3 digits of model number.
(See nameplate on side of antenna.)
_____________
Antenna
What is maximum nozzle length for
Extension
which the antenna can be used?
Select 8th digit of antenna model number.
(See nameplate on side of antenna.)
_____________
Antenna
Is the antenna mounting NPT, BSP,
Mount
or flanged?
_____________
Heat
Is there a heat extension connected
Extension
to the antenna?
_____________
Tank Height
What is the tank height?
_____________
Stillwell ID
What is the Inner Diameter (ID).
Enter 0 if not applicable.
_____________
Dielectric
What is the dielectric of the process
Range
medium?
_____________
Turbulence
What amount of turbulence is expected?
_____________
Foam
What amount of foam is expected?
_____________
Rate of
What is the expected maximum rate
Change
of level change?
_____________
Primary
Select Level, Volume, or Flow
Variable
_____________
4 mA
What is the 0% reference point for the
Setpoint
4.0 mA value?
_____________
(LRV)
20 mA
What is the 100% reference point for
Setpoint
the 20.0 mA value?
_____________
(URV)
PV Alarm
What output current is desired when a
Selection
failure indicator is present?
_____________
Damping
How much damping (averaging) is
required? Default = 1 second