21
48-635 E3 Modulevel
®
Displacer Level Transmitter
2.6.5.2
Measurement Type: Interface Level (cont.)
Display
Action
Comment
24
History
(current status)
Diagnostic Display to view
present status and recent
exceptions
Press
to view recent exceptions, up to ten events
25
Run Time
xxxx.x h
Diagnostic Display showing
elapsed time since power on
or History Reset
26
History
Reset
Factory only
27
MeasType
(select)
Factory set
Level, IfcLevel, or Density
28
Model
(select)
Factory set
E3A, E3B, E3C, E3D, E3E, E3F
29
SpringSG
(select)
Factory set
Model specific
30
SprgRate
x.x
Factory set
Model specific
31
SprgMatl
Factory set
Model specific
32
TempLimt
xxx F
Factory set
Model specific—maximum process temperature for which unit is
suitable
33
Length
xx.xx lu
Factory set
Model specific—length of measuring range
34
Diameter
x.xxx in
Factory set
Model specific—outside diameter of displacer
35
Weight
xx.x oz
Factory set
Model specific—weight of displacer and stem assembly
36
Lower SG
x.xx
Factory set
37
Upper SG
x.xx
Factory set
38
CalSelct
(select)
Select Factory or User
Calibration
Selects Calibration Parameters used to calculate the
measured PV
39a
Factory
Cal Menu
Press
to display the
Factory Calibration sub-menu
CalSelct = Factory
Submenu on page 22
39b
User
Cal Menu
Press
to display the User
Calibration sub-menu
CalSelct = User
Submenu on page 22
40
AdjSnrLo
Diagnostic Display
41
AdjSnrHi
Diagnostic Display
42
Conv Fct
xxxx
Diagnostic Display
43
Scl Ofst
xxx
Diagnostic Display
44
LVDT%
xx.xx %
Diagnostic Display
45
Chan 0
Diagnostic Display
46
Chan 1
Diagnostic Display
47
NSP Value
Diagnostic Display
48
ElecTemp
xxx F
Diagnostic Display
Present temperature in electronics compartment
49
Max Temp
xxx F
Diagnostic Display
Maximum electronics temperature recorded
50
Min Temp
xxx F
Diagnostic Display
Minimum electronics temperature recorded