PRONA/supination

-10     STOP    5   10

Please wait

KINETEC

Maestra V01.0 

Language:

english

Language:

french

Ok

Switch on/off

 

Use of the hand control 

 

Select your language 

 

Beginning 

Keys to press 

Display 

Remarks 

Switch the unit ON 
 

 

 
 
 
 
 
 

Check if the hand control is not locked  
(See page 5). 

Press the 2 keys in the same time  

 

 

The display indicate the language selected. 

To change the language  

or 

 

The English language is selected. 
 
English 
French  
German  
Italian  
Spanish 
 

To validate the new language. 

 

 

To exit and confirm the new language, switch 
OFF and Switch ON the unit. 

 

Locking the hand control setting 

 

The hand control allows the patient to control the machine as appropriate. 

 

Simultaneously press the                and                keys to lock the hand control,  

 

The display reads "LOCK", you can not change the parameters, if you try the display reads "LOCK". 

 

To unlock the hand control, simultaneously press the                and                keys, The display reads "UNLOCK”. 

 

 

We recommend that you lock the hand control when you give it to the patient. 

 

Comment: The hand control is unlocked when you switch the unit ON/OFF. 

 

START/STOP/REVERSE function 

 

As with all KINETEC systems, KINETEC Maestra is equipped with a START/STOP/REVERSE function. 

When the unit is running, the display reads RUN  

 

Press the                         key of the hand control. The movement stops. The display reads STOP. 

 

Press the                        key of the hand control. The movement starts in the opposite direction and the display reads RUN. 

 

 

 

Caution: 

For optimum safety, always give the hand control to the patient before starting the system. 

 

Содержание 8080

Страница 1: ...User manual Before use please read this document Kinetec reserves the right to effect technical modifications EN 467896276 08 2011 MAESTRA...

Страница 2: ...____________________________________________ 15 Thumb opposition __________________________________________________ 16 Complete fist _____________________________________________________ 17 Intrinsic...

Страница 3: ...ulotomies arthrolysis and tenolysis for post traumatic stiffness of MCP PIP and SIP joints Synovectomies Clinical Benefits Breaks the cycle of trauma inflammation and the loss of range of motion Preve...

Страница 4: ...Spare parts 10 Miscellaneous textiles 11 Ulnar deviation mobilization accessories 12 Power supply cable transport position 13 Hand control 14 Liquid crystal display 2 lines of 16 characters 15 Select...

Страница 5: ...not touch the moving parts while the unit is running pinching risk Electrical Connection KINETEC Maestra is a type B class I device Before connecting the device to the power supply check that the main...

Страница 6: ...hine as appropriate Simultaneously press the and keys to lock the hand control The display reads LOCK you can not change the parameters if you try the display reads LOCK To unlock the hand control sim...

Страница 7: ...press Lower limit 90 50 50 30 0 30 0 0 0 Upper limit 90 90 140 60 180 225 90 180 70 Speed 1 to 9 from 150 to 440 per minute Force 1 to 6 Extension Pause 0 to 900 seconds 15 minutes Flexion Pause 0 to...

Страница 8: ...pause in extension or flexion limit Successive presses on this key selects the pause at the extension or flexion limit To change the pause if necessary or The new pause value blinks To validate the ne...

Страница 9: ...record the program 10 Then The program 10 has been recorded and the display indicates the next program so you can change another program OR To cancel the program The program 10 has been cancelled and...

Страница 10: ...omments When a program has been deleted the display shows Program table Timer Pause on lower limit Pause on upper limit Force Speed Upper Limit Lower limit Movement type Program number 1 2 3 4 5 6 7 8...

Страница 11: ...eck if the hand control is not locked See page 5 To access the program mode The program number blinks To change the program if necessary or To change the program if necessary To read the speed value D...

Страница 12: ...address the pain threshold of a patient during a session IMPORTANT Can be used only when the machine RUN Beginning Keys to press Display Remarks The unit is running Check if the hand control is not lo...

Страница 13: ...lexion The joints also permit the forearm support to be tilted up or down However if this is done the angle of tilt should be taken into account when setting the range of motion on the display Setting...

Страница 14: ...er fit individual hand lengths and polar radii depending on the fingers or thumb to be mobilized the guide will be oriented as follows by unscrewing the nut 8 and lifting the guide 7 in position A or...

Страница 15: ...pination Choice of the Pronation Supination movement Press and adjust your parameters see page 7 Or select a program see page 10 Extension Flexion of the wrist The KINETEC Maestra provides a motion fr...

Страница 16: ...ion from 50 to 140 Choice of the 2nd movement of the wrist Press 2 times and adjust your parameters see page 7 Or select a program see page 10 Ulnar deviation The KINETEC Maestra provides a motion fro...

Страница 17: ...de with the screw 1 Remove the guide 2 from Set up the extension 3 of guide Set up the driving mechanism the driving mechanism Set up the guide 2 on the motor hub see p 12 for more details Set up the...

Страница 18: ...o 225 Choice of the complete fist movement Press and adjust your parameters see page 7 Or select a program see page 10 Intrinsic The KINETEC Maestra provides a motion from 0 from 90 Choice of the intr...

Страница 19: ...ment Press 2 times and adjust your parameters see page 7 Or select a program see page 10 Selective mobilization of the DIP joint The KINETEC Maestra provides a motion from 0 to 70 Choice of the Select...

Страница 20: ...100 240 V Frequency 50 60 Hz Power consumption 50 VA Device of type B class I IP 20 Fuse T 750mA 250V 6 3x32mm KINETEC order 4610007434 Environment Storage transport conditions Temperature 40 to 160...

Страница 21: ...t of the component or components found to be defective KINETEC guarantees its continuous passive motion systems for 2 years against all defects of manufacture from the date of purchase by the consumer...

Страница 22: ...______________ Technician ____________________________________________ _________________________________________ Running time see page 18 _________________________________ ____________________________...

Отзывы: