Data Logger Manual
Page 3
Section 1: Device Safety
CAUTION:
DEVICE MUST BE USED ONLY IN A MANNER CONSIS-
TENT WITH THIS MANUAL.
NOTICE:
WHEN 230VAC SUPPLY USED, MADGETECH SPECIFIES
USE OF AC SUPPLY PART # T35-9-00R-3 MANUFACTURED BY
ENG ELECTRIC. THIS SUPPLY IS AVAILABLE FROM MADGETECH.
1.1 Model information
•
Model: PRHTemp2000
• Description: Measure & record temperature, humidity, atmospheric
pressure with LCD display
• Manufactured by MadgeTech, Inc. Warner, NH USA
1.2 Specifications
Temperature
Range: -20 to +60ºC
Resolution: 0.ºC
Calibrated Accuracy: ±0.5ºC (0 to +50ºC)
Humidity
Range: 0 to 95%RH
Resolution: 0.%RH
Calibrated Accuracy: ±3%RH (±2%RH typical at 25ºC)
Specified Accuracy Range: +0 to +40°C, 0 to 80%RH
Pressure
Range: 0 to 30PSIA
Resolution: 0.002PSIA
Calibrated Accuracy: ±.0%FSR at 25ºC; ±0.2% typical
Dot-Matrix LCD
Dimensions: 2.5” x .375” (63mm x 35mm)
Text: Configurable channel text size
Indicators: Power, status, memory
Backlight: Configurable w/auto shut-off and contrast adjustment