Macurco CD-xx H/MC Manual
REV
–
1.1
[34-2900-0510-8]
29
|
P a g e
14.
If the calibration fails, the display will show “CAL6” for 1 second and then change to “Fail.3” for 30 seconds before
changing back to CAL for 30 seconds and then displays “t00/t020”
42T
Note:
42T
this Trouble code will only be cleared by a successful calibration of the sensor.
If this occurs, check the pressure gauge on the regulator. If the pressure is less than 25 psi the flow of gas may not be
adequate to properly calibrate the unit. If there is proper pressure in the cylinder repeat steps 5 through 12. If the unit
fails to calibrate twice contact Technical Support: 1-844-325-3050.
42T
Note:
42T
It is critical to confirm to use 400 ppm CO
P
2
P
gas calibration be used.
15.
Once the calibration has passed, remove gas and disassemble the cylinder and regulator.
16.
Re-assemble the CD-6MC (make sure the LED is aligned with the front case hole).
17.
See Calibration Flowchart on the inside of the housing, or section
Содержание CD Series
Страница 31: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 31 P a g e 9 Appendix B Menu Structure 9 1 Main Menu ...
Страница 32: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 32 P a g e 9 2 Auto Test Menu bUZ ...
Страница 33: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 33 P a g e 9 3 Configuration Menu CON ...
Страница 34: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 34 P a g e ...
Страница 35: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 35 P a g e ...
Страница 36: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 36 P a g e ...
Страница 37: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 37 P a g e ...
Страница 38: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 38 P a g e ...
Страница 39: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 39 P a g e 9 4 Select Test Menu tst ...
Страница 40: ...Macurco CD xx H MC Manual REV 1 1 34 2900 0510 8 40 P a g e 9 5 CAL Menu CD xxMC Only ...