Sensopress_LCD_Rev_00_Eng.doc Eng
17
5.1
CALIB Mode
It is possible to enter the CALIB mode from any other mode by selecting the sign
GO to CALIB
with
the
Menu
button and by pushing
Yes/+
.
In the
CALIB
mode it is possible to save the reading realized by the sensor that has been put through
level zero and through the maximum level that can be fit by the user.
This makes the reading correct in all the expected measure field.
When entered this, there is the visualization of Figure 14.
By pushing the
No/-
button it is possible to enter the procedure of zero setting.
By pushing the
Yes/+
button it is possible to enter on the procedure of the setting of the maximum
level.
Figure 14
5.1.1
Zero Calibration
It is possible to enter the zero calibration by pushing the
No/-
button after entered the
CALIB
.
On the display it is visualised the screen of Figure 15.
Figure 15
Make sure the sensor is at zero level.
Push the
Menu
button to enter the zero setting and it is visualized the screen of Figure 16.
Push button No/– to calibrate the zero.
Push button Yes/+ to calibrtae the high level.