Instruction Manual
MAN-004
7
www.macmedical.com
Installation
1. Carefully uncrate the scrub sink. The mounting hardware is under the sink, secured to the shipping materials.
Be sure to remove all hardware before discarding packaging.
2. Check water supply and waste terminal locations -
A.
Be sure all connections are safely accessible to the sink and in proper working condition. NOTE: For
proper sink operation the hot water temperature must be 120° F.
B.
Install sink near a hot water source.
3. Locate Infrared sensor activated sinks near a 110/120V outlet power source.
4. Attach wall mount sink support brackets to the wall.
A.
“Z” Brackets for mounting the sink to the wall come with the sink (Fig.
5). Recommended mounting specifications for Z brackets shown in
Fig. 13 (SS32), Fig. 16 (SS64), and Fig. 19 (SS96) on page 9 through
page 11.
B.
IMPORTANT:
For wall mounted units, the wall structure must be
capable of supporting the load.
C. Optional wall mounting systems (in-wall chair carrier and pedestal
style) and their mounting specifications shown in Fig. 15 (part S0001 for
sink SS32), Fig. 18 (part S0002 for sink SS64), and Fig. 21 (part S0003 for
sink SS96) on page 9 through page 11.
5. Install the soap spout on the sink (See page 8).
6.
Mount the sink body onto the Z Brackets that were attached to the wall.
A. With the help of an assistant, slide the sink over the “Z” brackets until it
locks securely into place (Fig. 6 - Fig. 7, Z brackets shown in green).
7. Connect the drain pipe to the drain
8. Connect the water supply to the plumbing (page 12).
Fig. 5: Z brackets
Fig. 6: Z brackets on sink
Fig. 7: Cross section showing Z bracket on sink