Instruction Manual
MAN-027
53
www.macmedical.com
4. Approaching a stored position
5. Setting the shown height on the display
6. Reset of the Control Box
A. Redefine end positions («S 7»)
B. Reset control box to factory settings («S 0»)
This function is designed to approach a stored position.
Press one of the buttons
1
2
3
4
.
The system approaches and stops at the stored position.
→
The displayed height can be adjusted with this feature.
1.
Drive to any desired height and press the button
S
(Save).
Display:
2.
Keep the button
pressed for about 5 seconds, until the display starts flashing.
Example:
3.
Now the button
(plus) or
(minus) can be used to set the current height.
While doing so, the system does not move!
4.
With the correctly set value the new height is saved by pressing
S
(Save).
1.
Press the buttons
1
,
2
and
simultaneously, until «S 5» or «S 7» is displayed.
2.
Press the button
until «S 7» appears on the display.
3.
Press the button
S
(Save).
The display is flashing «068»
carry out initial operation according to chapter 4.
1.
Press the buttons
1
,
2
and
simultaneously, until «S 5» or «S 7» is displayed.
2.
Press the button
until «S 0» appears on the display.
3.
Press the button
S
(Save).
The display is flashing «068»
carry out initial operation according to chapter 4.