background image

48NDT

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201  Fax: +81(6)6659-8510  E-mail: [email protected]

EM-9444-A  Rev.9  P. 2 / 8

 

AND ....

• The unit is designed to function as soon as power is sup-

plied, however, a warm up for 10 minutes is required for 
satisfying complete performance described in the data 
sheet.

COMPONENT IDENTIFICATION

CJC

Sensor

50

40

30

20

10

0

10

20

30

40

50

m

3

/h

%

mA

100 SET

H

S

P

Z

L

90

80

70

60

50

40

30

20

10

0

M

Brackets

Body

A

Scale Plate

Pull up the part ‘A’ when replacing the scale plate.

INSTALLATION

 

PANEL CUTOUT  unit: mm

138

+1 –0

31.5

Note 1:  Observe at the minimum of 3 cm above and below the units 
              for heat dissipation.
Note 2:  No bezel is needed when the panel is cut according to the 
              above drawings.

Panel thickness: 1.6 – 8.0 mm

■ 

SINGLE MOUNTING 

    (Conform to degree of protection IP65)

138

+1 –0

L

Panel thickness: 1.6 – 8.0 mm

 
L = {31.5 + 36 × (N – 1)}
(N : number of units)

+1

–0

■ 

CLUSTERED MOUNTING 

    (Not conform to degree of protection IP65)

+1

–0

61.5 min.

 

USING BEZELS

Bezels are used when the unit needs to be adapted to an ex-
isting panel cutout.  It is not required when the panel cutout 
size is as specified in the unit’s data sheet (31.5 × 138 mm).
Three types of bezels are selectable when ordering.
  Standard type:   Fits into M-System’s 48 Series’ panel   

cutout (38 × 139.5 mm)

  Option /D type:  Fits into DIN panel cutout
     

 

 

 

(33 × 138 mm)

  Option /F type:  Fits into Fuji Electric PAJ, PAK, PBA,
         

 

 

etc, panel cutout (44 × 138 mm)

Slide the top and bottom parts of the bezel into the back of 
the front cover as shown below.

Bezel

When multiple units are to be clustered side by side, insert 
one bezel between two units as shown below.

• When mounting two or more indicators side by side:

 

HOW TO MOUNT THE UNIT ON A PANEL

1) Insert the unit into the panel cutout.  The watertight 

packing must be in place to hold the unit.  Do not remove 

it even when the protection is not required.

  The IP65 front cover protection is effective only with sin-

gle mounting, with the designated cutout size (31.5 ×138 

mm).

Panel

Watertight Packing

Содержание 48NDT

Страница 1: ...fect the protection level of this unit when it is integrated in a panel system The user may have to review the CE requirements in regard to the whole system and employ additional protective measures t...

Страница 2: ...CLUSTERED MOUNTING Not conform to degree of protection IP65 1 0 61 5 min USING BEZELS Bezels are used when the unit needs to be adapted to an ex isting panel cutout It is not required when the panel...

Страница 3: ...to the connection diagram on the side of the unit Attach the CJC sensor together with input wiring to the input terminals The CJC sensor is not interchangeable Do not remove No 2 terminal screw so th...

Страница 4: ...4 9 19 136 5 5 37 144 5 67 8 5 33 10 4 41 7 5 30 136 5 5 37 144 5 67 OPTION F BEZEL 4 CONNECTION DIAGRAM T C comp leadwire T C comp leadwire T C comp leadwire 12 13 14 15 16 12 13 14 15 16 LLa 11 La...

Страница 5: ...orange green green red red red H L 50 40 30 20 10 0 10 20 30 40 50 100 SET H S P Z L 90 80 70 60 50 40 30 20 10 0 SET ENTER Button H Setpoint Adj Alarm Function Span Adj Setting Decimal Point Position...

Страница 6: ...lay for L Setpoint Increase Decrease Complete Digital Indicator SET L LL M Press at once ON Time Delay for LL Setpoint Increase Decrease Complete Digital Indicator SET LL Z M Press at once Zero Adjust...

Страница 7: ...inuous ON state 4 Press UP or DOWN buttons until the bargraph shows 0 and the digital indicator comes to a desired value and press SET DIGITAL INDICATOR ZERO ADJUSTMENT 1 Hold down M for 3 seconds and...

Страница 8: ...1 minute after it has been used to choose a indicator LED programming mode During this period pressing the M button briefly is enough to choose another indicator After 1 minute it turns to deac tivat...

Отзывы: