47D Series Modbus Reference Guide EM-9501-C Rev.4
3.5 INPUT SETTING
3.5.1 COMMON SETTING
ADDR. WORD CONTENTS
UNIT
1001
1
Averaging type
0 : Simple average
1 : Moving average (*)
1002
1
Averaging time
0 : No averaging (*)
1 : 2 samples
2 : 4 samples
3 : 8 samples
4 : 16 samples
5 : 32 samples
6 : 64 samples
7 : 128 samples
8 : 256 samples
9 : 512 samples
1003
1
High-pass filter
0 : High-pass filter OFF (*)
1 : High-pass filter ON
1011
1
Low-end cutout
0 : Low-end cutout OFF (*)
1 : Low-end cutout ON
2 : Low-end cutout by absolute value
1012
1
Low-end
cutout
value
0 ... 999
V M AC
Decimal point position setting
Factory setting : 0
T R
°C/100 or °F/100
3.5.2 FORCED ZERO, TARE ADJUSTMENT
V M
AC
ADDR. WORD CONTENTS
1021
1
Forced zero
0 : Disable Forced zero (*)
1 : Enable Forced zero
2 : Enable Tare adjustment
1022
2
Force zero value
Set value : -20000 ... 100000
1024
2
Tare adjustment value
Set value : -20000 ... 100000