- 28 - 

9.2.4.2

 

Fill a small amount (0,5 ml till maximum 2,0 ml) of distilled or sterile water into 
the medication chamber and close the M-neb

®

 dose

+

 nebulizer unit MN-300/8, 

by replacing the screw cap and closing it by rotating clockwise 80°. 

 

9.2.4.3 

After this, connect the filled M-neb

®

 dose

+

 nebulizer unit MN-300/8 with the M-

neb

®

 dose

+

 mesh nebulizer MN-300/8 (control unit). To do this, connect the 

female USB socket of the control unit with the male USB connector of the 
nebulizer unit. 

 

9.2.4.4 

Switch the M-neb

®

 dose

+

 mesh nebulizer MN-300/8 (control unit) on. For this, 

press (> 1 second) the on/off button of the M-neb

®

 dose

+

 mesh nebulizer MN-

300/8 (control unit). The M-neb

®

 dose

+

 mesh nebulizer MN-300/8 switches on 

automatically. As indication, you will hear a short acoustic signal tone and the 
battery status indicator continuously lights up green. 
The operational readiness of the M-neb

®

 dose

+

 mesh nebulizer MN-300/8 

(control unit) is established after the operating mode status indicator has 
flashed three times blue briefly. 
The M-neb

®

 dose

+

 mesh nebulizer MN-300/8 is now in triggered mode. 

 

9.2.4.5 

Now you can change the operating mode from the triggered mode to continuous 
mode  in  a  time  window  of  60  seconds,  which  is  equivalent  to  the  maximum 
standby time. For this, press (< 1 second) the on/off button of the M-neb

®

 dose

+

 

mesh nebulizer MN-300/8 (control unit). Two consecutive, short, acoustic signal 
tones sound and the operating mode status indicator lights up continuously blue. 

 

9.2.4.6 

The  M-neb

®

  dose

+

  mesh  nebulizer  MN-300/8  is  now  in  the  continuous  mode 

and  continuously  nebulises  the  filled  distilled  or  sterile  water.  Once  the  filled 
distilled  or  sterile  water  has  been  completely  nebulised,  the  auto  off  function 
automatically switches off the M-neb

®

 dose

+

 mesh nebulizer MN-300/8. 

 

WARNING 

Do not inhale the aerosolised distilled or sterile water. 
Ensure sufficient ventilation of the room in which you perform the cleaning. 
 

9.2.4.7

 

Now  disconnect  the  M-neb

®

  dose

+

  nebulizer  unit  MN-300/8  from  the  M-neb

®

 

dose

+

 mesh nebulizer MN-300/8. Disconnect the screw cap of the M-neb

®

 dose

+

 

nebulizer unit MN-300/8 and put the M-neb

®

 dose

+

 nebulizer unit MN-300/8 on 

a dry, absorbent surface to dry. 

 

9.2.4.8

 

After the  M-neb

®

 dose

+

 nebulizer unit  MN-300/8 has dried in the air, you can 

reconnect  the  screw  cap  of  the  M-neb

®

  dose

+

  nebulizer  unit  MN-300/8. 

Reconnect  the  M-neb

®

  dose

+

  mouthpiece  MN-300/8  again  with  the  M-neb

®

 

dose

+

 nebulizer unit MN-300/8. 

 
 

10.0  Maintenance 

The M-neb

®

 dose

+

 mesh nebulizer MN-300/8 should be checked after a maximum of 24 

months and subjected to a technical safety inspection. All repairs or maintenance may only 
be done by a dealer qualified by NEBU-TEC med. Products Eike Kern GmbH. 
 

Содержание Dose+

Страница 1: ...ultrasonic nebulizer Instructions for use M neb dose mesh nebulizer MN 300 8...

Страница 2: ...2...

Страница 3: ...TEC med Produkte Eike Kern GmbH Kreuzfeldring 17 63820 Elsenfeld GERMANY Tel 49 0 6022 610 62 0 Fax 49 0 6022 610 62 99 Email info nebu tec de Web http www nebu tec de M neb dose mesh nebulizer MN 30...

Страница 4: ...out inhalation 17 6 4 Switch off M neb dose mesh nebulizer MN 300 8 19 6 5 Changing the operating modes 20 6 6 Use of the continuous mode 22 7 0 Power supply 23 8 0 Charging of the M neb mesh nebuliz...

Страница 5: ...Device of protection class II Type B applied part 0197 CE mark including designation of the identification number of the notified body which has performed the conformity assessment procedure Follow in...

Страница 6: ...is damaged or opened Keep away from sunlight designates a medical device that needs protection from light sources Store in a dry place designates a medical device that must be protected against moistu...

Страница 7: ...Important information is highlighted by the following expressions WARNING Important safety information on hazards that can cause personal injury ATTENTION Important information on operation steps that...

Страница 8: ...bserved to verify its proper operation in the configuration used 15 Do not open the housing All repairs or maintenance may only be done by a dealer qualified by NEBU TEC med Produkte Eike Kern GmbH Fa...

Страница 9: ...se any other commercially available USB cable with the USB connection of the M neb dose mesh nebulizer MN 300 8 control unit REF 20100262 CAUTION 1 Do not touch the aerosol generators in the M neb dos...

Страница 10: ...vices without implantable and infectious products Therefore the electronic components of the M neb dose mesh nebulizer MN 300 8 M neb dose mesh nebulizer MN 300 8 control unit and M neb power cord MN...

Страница 11: ...s two modes of nebulisation triggered mode and continuous mode The basic setting of the M neb dose mesh nebulizer MN 300 8 is the triggered mode Any aerosol production of the M neb dose mesh nebulizer...

Страница 12: ...nuous mode is selected in the following chapter 6 5 Changing the operating modes 3 3 Automatic switch off Regardless of the nebulisation mode the M neb dose mesh nebulizer MN 300 8 switches off automa...

Страница 13: ...M41076 0605 A 2 pc 20100268 M neb dose nebulizer unit MN 300 8 incl M neb dose mouthpiece MN 300 8 1 pc 20100207 M neb Etui MN 300 X NOTE Before the first use of the M neb dose mesh nebulizer MN 300 8...

Страница 14: ...the accessories All components of the M neb dose mesh nebulizer MN 300 8 are listed in the table below Number Designation Figure 20100262 M neb dose mesh nebulizer MN 300 8 control unit 20100206 M neb...

Страница 15: ...illing the M neb dose nebulizer unit MN 300 8 6 1 1 Open screw cap using an 80 rotation counter clockwise then lift the cap upwards 6 1 2 Pour medication into the medication chamber 6 1 3 Place screw...

Страница 16: ...mesh nebulizer MN 300 8 control unit After this connect the filled M neb dose nebulizer unit MN 300 8 with the M neb dose mesh nebulizer MN 300 8 control unit To do this connect the female USB socket...

Страница 17: ...green 6 3 2 The operational readiness of the M neb dose mesh nebulizer MN 300 8 is established after the operating mode status indicator has flashed three times blue briefly 6 3 3 The M neb dose mesh...

Страница 18: ...around the mouthpiece of the M neb dose mesh nebulizer MN 300 8 3 again and inhale through your mouth To exhale put down the mouthpiece of the M neb dose mesh nebulizer MN 300 8 again and do not exha...

Страница 19: ...d the M neb dose mesh nebulizer MN 300 8 must be switched off Press 1 second the on off button of the M neb dose mesh nebulizer MN 300 8 control unit The M neb dose mesh nebulizer MN 300 8 switches of...

Страница 20: ...eadiness of the M neb dose mesh nebulizer MN 300 8 is not established and the device is in the switched off state 6 5 1 Press 1 second the on off button of the M neb dose mesh nebulizer MN 300 8 contr...

Страница 21: ...ator lights up again three times briefly 6 5 7 In order to switch off the M neb dose mesh nebulizer MN 300 8 press 1 second the on off button of the M neb dose mesh nebulizer MN 300 8 control unit A s...

Страница 22: ...The M neb dose mesh nebulizer MN 300 8 switches on automatically As indication you will hear a short acoustic signal tone and the battery status indicator continuously lights up green 6 6 2 The operat...

Страница 23: ...control unit may only be charged via the M neb power cord MN 300 X type GTM41076 0605 A The M neb dose mesh nebulizer MN 300 8 control unit may only be operated via the M neb power cord MN 300 X type...

Страница 24: ...24 8 0 1 Connect the supplied M neb power cord MN 300 X type GTM41076 0605 A for charging first with the M neb dose mesh nebulizer MN 300 8 control unit...

Страница 25: ...8 3 is charging the battery indicator light flashes green ratio 1 second green 1 second pause 8 0 5 If the M neb dose mesh nebulizer MN 300 8 has completed charging the battery indicator light flashes...

Страница 26: ...26 ATTENTION The charging of the internal rechargeable battery lithium ion battery is carried out exclusively with the supplied M neb power cord MN 300 X type GTM41076 0605 A...

Страница 27: ...dose mesh nebulizer MN 300 8 ensure that the M neb power cord MN 300 X type GTM41076 0605 A has been disconnected Never immerse the housing of the M neb dose mesh nebulizer MN 300 8 in water or a cle...

Страница 28: ...ndby time For this press 1 second the on off button of the M neb dose mesh nebulizer MN 300 8 control unit Two consecutive short acoustic signal tones sound and the operating mode status indicator lig...

Страница 29: ...ot lit when charging is completed Operating mode status indicator The operating mode status indicator LED blue lights blue three times briefly when the boot phase is terminated blue once briefly when...

Страница 30: ...ied in the category 8 medical devices without implantable and infectious products Therefore the electronic components of the M neb dose mesh nebulizer MN 300 8 M neb dose mesh nebulizer MN 300 8 contr...

Страница 31: ...ata Aerosol dispensing and particle size in accordance with DIN EN 13544 1 Annex CC were determined and established for the M neb dose mesh nebulizer MN 300 8 16 1 Performance data of the M neb dose m...

Страница 32: ...m the manufacturer or dealer for your M neb dose mesh nebulizer MN 300 8 Item number Designation Quantity 20100008 M neb dose Starter Kit MN 300 8 1 pc 20100104 M neb dose consumption material for a p...

Страница 33: ...t it is used in such an environment Immunity tests IEC 60601 test level Compliance level Electromagnetic environment guidelines ELECTROSTATIC DISCHARGE ESD according to IEC 61000 4 2 6 kV contact disc...

Страница 34: ...Veff D 1 2 P Radiated RF disturbance according to IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m D 1 2 P 80 MHz to 800 MHz D 2 3 P 800 MHz to 2 5 GHz Where P is the rated output of the transmitter in wat...

Страница 35: ...ansmitter W Safety distance dependent on transmission frequency 150 kHz to 80 MHz D 1 2 P 80 MHz to 800 MHz D 1 2 P 800 MHz to 2 5 GHz D 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3...

Страница 36: ...LGA Products GmbH Tillystra e 2 90431 N rnberg Germany ID 0197 Conformity assessment procedure EC Directive 93 42 EEC Annex II Section 3 EC certificate HD 60101825 0001 CE mark since 2016 23 0 Warrant...

Страница 37: ...37 Your notes...

Страница 38: ...38...

Страница 39: ...39...

Страница 40: ...NEBU TEC med Produkte Eike Kern GmbH Kreuzfeldring 17 63820 Elsenfeld Germany...

Отзывы: