21.0 Information on electromagnetic compatibility
Guidelines and manufacturer’s declaration – electromagnetic transmissions
The nebulizer MN-300/8 is intended for use in the ELECTROMAGNETIC ENVIRONMENT specified below.
The customer or the user of the nebulizer MN-300/8 should ensure that it is used in such an environment.
Electromagnetic interference
measurements
Conformity
ELECTROMAGNETIC ENVIRONMENT-
guidelines
RF emissions according to
CISPR 11
Group 1
The nebulizer MN-300/8 uses RF energy only
for its internal FUNCTION. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions according to
CISPR 11
Class B
The nebulizer MN-300/8 is suitable for use in
all establishments, including those in the living
area and those that are directly connected to
the PUBLIC SUPPLY NETWORK that supplies
buildings used for residential purposes.
Emissions of harmonics
according to IEC 61000-3-2
Class A
Emissions of voltage
fluctuations/flicker according to
IEC 61000-3-3
Complies
WARNING
The M-neb
®
dose
+
mesh nebulizer MN-300/8 may not be placed directly next to other
devices. If necessary, the device should be observed to verify its proper operation in the
configuration used.
Guidelines and manufacturer’s declaration – Electromagnetic immunity
The nebulizer MN-300/8 is intended for use in the ELECTROMAGNETIC ENVIRONMENT specified
below
.
The customer or the user of the nebulizer MN-300/8 should ensure that it is used in such an
environment.
Immunity tests
IEC 60601 test level
Compliance level
Electromagnetic environment guidelines
ELECTROSTATIC
DISCHARGE (ESD)
according to IEC 61000-
4-2
+
6 kV contact discharge
+
8 kV air discharge
+
6 kV contact discharge
+
8 kV air discharge
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Rapid transient electrical
interference
bursts according to IEC
61000-4-4
+ 2 kV for mains lines
+ 1 kV for input and output
leads
+ 2 kV for mains lines
+ 1 kV for input and output
leads
The quality of the supply voltage should
correspond to that of a typical commercial or
hospital environment.
Surge voltage/surges
according to IEC
61000-4-5
+
1 kV phase-to-phase
voltage/phase conductor
+
2 kV phase to earth
+
1 kV phase-to-phase
voltage/phase conductor
+
2 kV phase to earth
The quality of the supply voltage should
correspond to that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
fluctuations in the
supply voltage
according to IEC
61000-4-11
< 5% U
T
(> 95% dip of the
U
T
) for ½ periods
40% U
T
(60% dip of the U
T
)
for 5 periods
70% U
T
(30% dip of the U
T
)
for 25 periods
< 5% U
T
(> 95% dip of the
U
T
) for 5 s
< 5% U
T
(> 95% dip of the
U
T
) for ½ periods
40% U
T
(60% dip of the
U
T
) for 5 periods
70% U
T
(30% dip of the
U
T
) for 25 periods
< 5% U
T
(> 95% dip of the
U
T
) for 5 s
The quality of the supply voltage should
correspond to that of a typical commercial or
hospital environment.
When the user of the nebulizer MN-300/8 requires
continued operation during interruptions of energy
supplies, it is recommended to power the
nebulizer MN-300/8 of an uninterruptible power
supply or a battery.