6. Responsability
We will supply the circuit diagram of the instrument, electric component list, drawing
annotation and calibration details according to the customer´s need for repair.
If there is any need for enquiry of relative information and relative service or some
questions, please contact with us directly.
7. Transportation
During the transportation, be careful to protect it from wetness, upside down and violent
vibration. The relative humidlty should be 10% to 90%, and enviroment temperature
-25ºC to 40ºC. This instrument should be stored in a well ventilated room without
corrosive gas where the relative humidlty should be 10% to 80% and enviroment
temperature -10ºC to 40 ºC.
If the assembled instrument should be moved or tranported in short distance; please
lock all the moveble parts. Move this instrument carefully with hands pushing or carrying
its table. If for long distance transportation, please repack it with original package.
8. Optional Accessories
(purchase in addition)
8.1 10% measuring eyepieces
Replace the common eyepieces withthis one to measure the length and angle.
29
Trouble
Slit closes
automatically
Fixation target
is off
Possible Cause
The fuse doesn’t comply
with the specification.
The slit width control knob
is too loose.
The output plug is loose.
Remedy
Replace it with a suitable
fuse.
Adjust the tightness of
the control.
Insert the output plug
firmly.