Luvion AngelSounds Скачать руководство пользователя страница 2

ENGLISH - USER MANUAL

AngelSounds - fetal doppler

1. On/Off/volume slider
2. Power indicator light
3. Headphone socket(two)
4. Sensor
5. Battery compartment cover

Accessories

Introduction

The AngelSounds fetal doppler is a single-hand device 
for detection of the fetal heart. The fetal heart may be 
detected from early pregnancy, and more easily from 
pregnancies greater than 12 weeks. Every AngelSounds 
is supplied with a dry-cell battery. A dry-cell battery must 
NOT be placed in a recharging device and must NOT be 
disposed of by burning. The AngelSounds fetal doppler 
operates at a nominal frequency of 3.3MHz, using Dop-
pler ultrasound. There is an ‘on/off/volume key(1)’, it’s easy 
to operate. The LED working indicator light (2) shows 
working condition. A headset and an audio line are avail-
able. The AngelSounds fetal doppler is lightweight and 
designed to fit comfortably in your hand.

Operation instruction

Insert the headset jack securely into the socket(3). Apply 
a small amount of acoustic coupling gel to the examina-
tion site. Or put some gel on the top of the sensor (4).

Switching on On/Off/volume key (1)
This key is used to select either high or low volume level, 
too. Hold the AngelSounds Fetal doppler. Plug in the 
earphones.

The AngelSounds fetal doppler is now ready for use.

Operation

Place the transducer on the examination site and move 
it around slowly until a good signal is obtained. This will 
sound like ‘galloping horses’. It may be necessary to angle 
the transducer slightly to obtain the optimum signal.

Use of audio line

Insert one jack of the audio line into the main unit’s 
headset socket (any of the two), put another plug into the 
recorder, then the fetal heart beat can be recorded by the 
recorder. Move the transducer to the pregnant woman’s 
left chest to record her heart beat and play the mother’s 
heart beat to the baby. After the baby is born, this can 
calm the baby down. To switch off, just turn the ‘On/Off/
Volume Slider’.

Battery fitting or replacement

How to fit or replace the battery on the AngelSounds 
fetal doppler:
Remove the rear battery compartment cover (5) carefully. 
Remove the old battery. Fit the new battery (IEC6F22 9V 
alkaline). Then close the battery cover.

• 

Use only the specified battery type and insert the 
battery according to polarity instruction (check the 
+ and – poles).

• 

Always dispose of empty batteries in accordance 
with regulations. Do not dispose together with 
household garbage.

New alkaline battery will give more than 500 one minute 
examinations. Battery should be removed if the unit is 
not in regular use. After use, the AngelSounds fetal dop-
pler should be dried thoroughly with a soft cloth. 

Operating condition:

Do NOT leave the AngelSounds fetal doppler exposed to 

direct sunlight. 

Operating temperature: 0-40 °C 

Operating humidity: 0% - 85%

Battery: IEC 6F22 9V alkaline

Safety check list:

The AngelSounds fetal doppler     Fetal Heart Detector is 

designed to comply with BS5724 part 1, IEC601-1,UL544 

and other international medical safety standards for bat-

tery-operated (internally powered) medical equipment.

Warning

•  Device is sealed and NOT user-serviceable. Device must 

be serviced only by authorized and qualified person-
nel to maintain safety, and reliability. Damage may 
result if the AngelSounds Fetal doppler is knocked or 
dropped.

Classification:

Type of protection against electric shock: Internally 

powered equipment. 

Degree of protection against electric shock: Type B

Type B protection means that this equipment will comply 

with:

EN 60601-1/ Medical Electrical Equipment Part 1:

IEC 60601-1 General Requirements for safety

EN 60601-1-2/ Standard for electromagnetic compat-

ibility

IEC 60601-1-2 requirements for medical electrical equip-

ment

U.S. Federal law restricts this device to use on or by the 

order of a Physician. Degree of protection against harmful 

ingress of water: Ordinary equipment Mode of operation: 

design for continuous operation

Degree of safety of application in the presence of a flam-

mable

Anaesthetic mixture with air or with oxygen or nitrous 

oxide:

Do not use in the presence of flammable anaesthetics.

This detector is not explosion-proof and must not be 

used in the presence of flammable anaesthetics.

Statement

The medical electrical equipment needs special precau-
tions
regarding EMC and needs to be installed and put into 
service according to the EMC information provided in 
the accompanying documents (this instruction). Portable 
and mobile RF communications equipment can affect the 
medical electrical equipment.
The equipment is without a manual sensitivity adjust-

ment, hence: The minimum amplitude or value of patient 
physiological signal is 90dB.

Warning:

•  Operation of the equipment or system below this am-

plitude or valuemay cause inaccurate results.

•  The use of accessories, transducers and cables other 

than those specified, with the exception of transducers 
and cables sold by the manufacturer of the equipment 
or system as replacement parts for internal compo-
nents, may result in increased emission or decreased.

Acoustic output data

The acoustic output of the AngelSounds fetal doppler 
transducer has been measured in water using a calibrated 
hydrophone at authorized Laboratory. Normalized values, 
which estimate the maximum ‘in-situ’ dosage to tissue at 
the point of highest intensity in the beam path have been 
calculated. These data are presented in the following 
tables.

Ultrasound safety considerations and data

General Diagnostic ultrasound has been in use for over 
25 years with no confirmed adverse effects on patients or 
instrument operators at the intensities typical of present 
diagnostic instruments. Although the total absence of ad-
verse effects to human subjects after such extensive use 
at diagnostic power levels is gratifying, available data are 
not conclusive and the possibility that biological effects 
may be identified in the future remains. It is therefore 
deemed desirable by medical and other scientific authori-
ties in this field that exposure to ultrasound should be 
limited to the duration and intensity appropriate for the 
clinical objective. Because fetal tissue could be more 
sensitive to biological effects by reason of pregnant 
subjects be kept to a minimum. At present, there is a clear 
consensus that the benefits to patients of prudent use of 
diagnostic ultrasound outweigh the risks, if any, that may 
be present.
AngelSounds fetal doppler is a portable battery operated 
detector designed to detect fetal life and confirmation of 
continued life during pregnancy. Minimizing patient ex-
posure Acoustic output of the AngelSounds fetal doppler 
is internally controlled and cannot be changed by the op-
erator in the course of the examination. The duration is, 
however, fully under the control of the operator. Mastery 
of the techniques described in the operating instructions 

2

3

Отзывы: