Appendix J:Precision and Accuracy
J.1
Precision
Precision is the ability of a methodology to repeatedly obtain the same value. Your
system, using DEXA, has been tested in the factory to ensure reliable measurements
(precision error <0.01 g/cm
2
). For comparison, a series of measurements over time
with conventional DPA typically shows a slightly higher variability of .02 to .03 g/
cm
2
. The measurements used a Gd
153
radionuclide source on a phantom simulating a
spine or femur.
The DPX has shown a better precision than Gd
153
based DPA (about .01 g/cm
2
error on
spine and femur phantoms). (Refer to tables J-1 and J-2.) Measurements with the
DPX system were repeated on a spine phantom and a femur phantom to assess
precision. AP spine measurements were made 40 times on 6 subjects over 1 week and
50 times in 1 subject over 1 month. The standard deviation (SD) was about .01 g/cm
2
for L2–L4, which corresponds to about 1% at an average density of 1 g/cm
2
. Femur
measurements were made 75 times on 15 subjects (5/case) over 1 week and 50 times
on 1 subject over 1 month. The precision error was about .015 g/cm
2
for the neck
region, or about 1.5% at a typical density of 1 g/cm
2
. These precision results were
summarized by Mazess et al (1989).
1
J
Precision and Accuracy
Содержание DPX-IQ
Страница 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Страница 12: ...xii Table of Contents DPX IQ Operator s Manual...
Страница 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Страница 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Страница 61: ...DPX IQ Operator s Manual AP Spine A 5 Figure A 5 AP Spine Analysis Report Figure A 6 AP Spine Ancillary Report...
Страница 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Страница 76: ...B 8 Femur DPX IQ Operator s Manual Figure B 8 Femur Analysis Report Figure B 9 Femur Ancillary Report...
Страница 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Страница 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Страница 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Страница 107: ...DPX IQ Operator s Manual Forearm E 7 Figure E 9 Forearm Analysis Report Figure E 10 Forearm Ancillary Report...
Страница 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Страница 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Страница 166: ...This page left blank intentionally...