DPX-IQ Operator’s Manual
Specifications
H-5
•
Shock and Vibration:
Make sure the scan table does not receive shock greater
than 1 G for more than 1 millisecond. Make sure the scan table does not receive
vibrations greater than 0.25 G at 5 Hz.
•
Temperature:
Make sure the temperature during system operation is 65°F–81°F
(18°C–27°C).
NOTE: When the system is turned off, or there is a power failure, you must
turn the system on and let it warm up for one hour. After one hour,
complete a Quality Assurance procedure.
H.5
Power Specifications
H.5.1
Leakage Current
Total System with Isolation Transformer: <100 microamperes.
Scan Table alone: <100 microamperes.
H.5.2
Scan Table
The following are the power specifications for the DPX product line. Check the power
specifications label on your scanner to determine which power specifications apply to
your scanner.
Bertan/Spellman
100 Range: 90-110 VAC 50/60 Hz 10A
P/N 0311 and 0312
115 Range: 103-127 VAC 50/60 Hz 10A
230/240 Range: 207-264 VAC 50/60 Hz 7.5A
H.6
X-ray Generator Specifications
H.6.1
X-ray Generator
Table H-1 contains information about the X-ray generator for DPX-IQ series devices.
Table H-1: X-ray generator technical information.
Classification
Class I Equipment
IEC 601-2-7 5.1
Degree of protection against electrical shock
Type B equipment
IEC 601-2-7 5.2
Protection against ingress of liquids
Ordinary medical electrical equipment
IEC 601-2-7 5.3
Connection to supply mains
Detachable power cord
IEC 601-2-7 6.1g)
Mode of operation
Continuous
IEC 601-2-7 6.1m)
Maximum X-ray tube voltage
76 Kv
IEC 601-2-7 6.1m)
Maximum X-ray tube current
5 mA
IEC 601-2-7 6.1m)
Rated mains voltage
100/115/230/240 volts
IEC 601-2-7 6.1j)1
Number of phases in mains
1
IEC 601-2-7 6.1j)2
Содержание DPX-IQ
Страница 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Страница 12: ...xii Table of Contents DPX IQ Operator s Manual...
Страница 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Страница 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Страница 61: ...DPX IQ Operator s Manual AP Spine A 5 Figure A 5 AP Spine Analysis Report Figure A 6 AP Spine Ancillary Report...
Страница 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Страница 76: ...B 8 Femur DPX IQ Operator s Manual Figure B 8 Femur Analysis Report Figure B 9 Femur Ancillary Report...
Страница 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Страница 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Страница 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Страница 107: ...DPX IQ Operator s Manual Forearm E 7 Figure E 9 Forearm Analysis Report Figure E 10 Forearm Ancillary Report...
Страница 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Страница 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Страница 166: ...This page left blank intentionally...