![Luna ODiSI 6 Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/luna/odisi-6/odisi-6_user-manual_1949788051.webp)
ODiSI 6
ODiSI 6
User’s Guide
Page 50
When a test is started, the specified port is opened and the ODiSI software begins listening
for a client connection. The client program can then initiate a TCP connection to the OMSP
server on the specified port number. As soon as a connection is established, the ODiSI begins
sending messages to the client. If the ODiSI loses the connection with the client, it resumes
listening for a new client connection.
Once a new connection is established, the ODiSI resumes sending data, but does not send
measurement data that was missed during the time the connection was interrupted.
The application data is one way. That is, the ODiSI sends data to the client but does not look
for any responses.
When a connection is made, the Status changes to “Client connected”. The Test Name and
Notes are included as Metadata to the client.
Remote Enabled
: Check to enable remote control of the ODiSI control software.
The Remote Control allows the user to connect to the ODiSI controller using a TCP/IP
connection and send Arm, Start, Stop, and Disarm commands to the instrument to control the
starting and stopping of measurements. See the section on
more information about the ODiSI Remote application and how to connect to an ODiSI
controller using the ODiSI Remote.
Note on Measurement Streaming and Saving:
There is an important distinction between
saved measurements and streamed measurements. Saved measurements are stored on the
instrument controller and are accessible for file playback. Streamed measurements are sent
to a third-party PC and no record of them is maintained by the instrument controller. You can
enable and utilize both of these features simultaneously.
Figure 6-29: Streaming Properties dialog