6 Error codes
The N32O meters have several diagnostic functions and settings built-in that allow to limit
the displaying. So the display may show and the status registers may store information
about the diagnosed error, event or fault. Possible messages and their potential causes
are listed below.
Message
Description
Overflow of the lower value of the measuring range value or the programmed
indication range.
Overflow of the upper value of the measuring range value or the programmed
indication range.
It is not possible to display the measuring value in the selected resolution - the
measurement result does not fit on the display. Lower the display resolution or
select the mode of automatic position of the decimal point.
Lost calibration. Please contact the technical support.
Lost real time clock settings. The message is displayed only when turning the
meter on. Time and date must be set. If the message still appears when turning
the meter on after setting the time and date, please contact the Service
Department because a backup battery may require to be replaced. This
message can be ignored if the clock settings are not significant in a given
application.
Measurement module error. It is not possible to make a measurement, please
contact the technical support.
Configuration data memory and calibration memory error. It is not possible to
use a meter, please contact the technical support.
No communication with the analog output module. Please contact the Service
Department.