N32O
User's manual
57
7508
7016
Maximum value of measuring quantity on the main input registered during the averaging window
(during the selected averaging period).
7509
7018
Measuring value on the auxiliary input without the mathematical functions, averaging using the
moving window method and without the individual characteristic.
7510
7020
Value additionally measured on the additional input.
7511
7022
Measuring value on the auxiliary input without the conversion, but averaged using the moving
window method.
7512
7024
Minimum value of the measured quantity on the auxiliary input (converted by the mathematical
functions, individual characteristics and averaged ).
7513
7026
Maximum value of the measured quantity on the auxiliary input (converted by the mathematical
functions, individual characteristics and averaged ).
7514
7028
Measuring value on the auxiliary input - measuring value and then averaged, converted by the
mathematical functions and converted in accordance with the configuration of the individual
characteristic.
7515
7030
Minimum value of measuring quantity on the auxiliary input registered during the averaging window
(during the selected averaging period).
7516
7032
Maximum value of measuring quantity on the auxiliary input registered during the averaging
window (during the selected averaging period).
7517
7034
Content of input no. 3 counter.
7518
7036
Content of input no. 4 counter.
7519
7038
Voltage of the backup battery.
7520
7040
CPU temperature.
7521
7042
Current time in the form of hh.mmss.
5.5.4.4 Registers 7600 – 7744 and 7200 – 7489
The 32-bit and the corresponding 16-bit registers with the configuration parameters.
Address (32-
bit float
registers)
Address
(value in 2 16-
bit registers)
Permissible values
Default
Description
Minimum and maximum displayed value
7600
7200
-99999...999999
-99999
Display narrowing lower threshold. If the value to be displayed is
below the threshold, the lower limit symbol is displayed.
7601
7202
-99999...999999
999999
Display narrowing upper threshold. If the value to be displayed is
above the threshold, the upper limit symbol is displayed.
Main and auxiliary input counter parameters
7602
7204
-99999...999999
1
CONSt parameter for the main input defines the value by
which the measurement result will be multiplied or
divided.
7603
7206
-99999...999999
999999
AUtO parameter for the main input. The value to which
the main input counter will count (for up counting) or the
initial value of the counter (for down counting).
7604
7208
-99999...999999
1
CONSt parameter for the additional input defines the
value by which the measurement result will be multiplied
or divided.
7605
7210
-99999...999999
999999
AUtO parameter for the additional input. The value to
which the main input counter will count (for up counting)
or the initial value of the counter (for down counting).
Analog output
7606
7212
-99999M...999999M
0
The quantity of the value controlling the analog output for
which the output will have the minimum value (according
to the output range).
7607
7214
-99999M...999999M
100
Quantity of value controlling the analog output for which