IS 12-TSP / IGA 12-TSP Manual
Introduction
•
9
2
Introduction
2.1
Appropriate use
IS 12-TSP and IGA 12-TSP are extremely precise and longer-term stable transfer standard
pyrometers which can be used for the checking of calibration sources in the respective
pyrometer spectral range.
2.2
Scope of delivery
Instrument with selectable optics, PC measurement and evaluation software „InfraWin“, works
certificate with 3 selected measuring points, user manual.
Note:
The connection cable is not included with the instrument and has to be
ordered separately (see
8
,
Reference numbers
).
2.3
Technical data
Measurement
Temperature Ranges:
IS 12-TSP:
530 ... 1900 °C
IGA 12-TSP:
200 ... 1020 °C
600 ... 2520 °C
250 ... 1400 °C
600 ... 3000 °C 300 ... 1700 °C
850 ... 2520 °C
Sub Range:
Any range adjustable within the temperature range
minimum span 51 °C
Signal Processing:
Photoelectric current, digitized
Spectral Range:
IS 12-TSP:
0.94 µm (at temperature ranges 530 to 1900 °C and
600 to 2520 °C / 3000 °C)
0.65 µm (at temperature range 850 to 2520 °C)
IGA 12-TSP:
1.57 µm
Resolution:
Up to 1000 °C: 0.1 °C at interface and display
Above 1000 °C: 0.1 °C at display, 0.01 °C at interface
< 0.025% of temperature range at the analog output
Emissivity
ε
:
0.100 to 1.000 in 1/1000 steps
Exposure Time t
90
:
1 ms (with dynamical adaptation at low signal levels); factory
setting 1 s, adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s
Measurement
Measurement Uncertainty *
(
ε
= 1, t
90
= 1 s, T
amb.
= 23 °C)
Below 1500 °C:
0.15% of measured value in °C + 1 °C
Above 1500 °C:
0.25% of measured value in °C
Above 2700 °C:
0.35% of measured value in °C
Repeatability:
(
ε
= 1, t
90
= 1 s, T
amb.
= 23 °C)
1 °C