IS 12-TSP / IGA 12-TSP Manual 

Introduction 

  11 

Electrical 

Power Supply: 

24 V DC (15 to 40 V DC) or 24 V AC (12 to 30 V AC), 48 to 62Hz 

Power Consumption: 

Max. 14 W  

Isolation: 

Power supply, digital interface, analog output are galvanically 

isolated against each other and housing 

 

*Additional measurement uncertainty due to offset drift of the signal converter at long 
temperature ranges. Tc = measuring temperature up to which an additional measurement 

uncertainty occurs when the ambient temperature differs from the reference temperature of  
23 °C 

Type 

MB/°C 

Tc/°C 

IS 12-TSP 

530 ... 1900  630 

IS 12-TSP 

600 … 2520  718 

IS 12-TSP 

850 … 2520  960 

IS 12-TSP 

600 … 3000  773 

IGA 12-TSP 

200 … 1020  265 

IGA 12-TSP 

250 … 1400  321 

IGA 12-TSP 

300 ... 1700  360 

 

 

Note: 

The calibration / adjustment of the instruments was carried out in accordance 

with VDI/VDE directive “Temperature measurement in industry, Radiation thermometry, 
Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4. 

For additional details on this directive, see http://info.lumasenseinc.com/calibration or 
order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany. 

 

* Note: 

  A surrounding temperature of 10 to 43 °C is permissible for measured 

temperatures between 10% and 100% of the temperature range for the stated 

measurement uncertainty. The measurement uncertainty increases by 1 °C for 
surrounding temperatures of between 0 to 10 °C or 43 to 60 °C. 

For temperature ranges of up to 5% of the measuring range an additional temperature 
drift of 0.2 °C/ °C (23 °C) should be added. The temperature drift is 0.05 °C/ °C (23 °C) for 
measured temperatures of 5% to 10% of the measuring range. 

 

Содержание IGA 12-TSP

Страница 1: ...MANUAL IMPAC Pyrometer IS 12 TSP IGA 12 TSP...

Страница 2: ...espective holders Service Centers LumaSense Technologies Inc North America Sales Service Santa Clara CA USA Ph 1 800 631 0176 Ph 1 408 727 1600 Fax 1 408 727 1677 LumaSense Technologies GmbH Other Tha...

Страница 3: ...for the connector on the back of the pyrometer 13 3 2 Connecting the pyrometer to a PC 14 3 2 1 Connecting to RS232 interface 14 3 2 2 Connecting to RS485 interface 15 3 3 Connection of additional ana...

Страница 4: ...s 25 4 12 kBaud baud rate 26 4 13 10 mA test current output 26 4 14 Wait time tw 26 4 15 Switch contact 26 5 Software InfraWin 27 5 1 Connecting the pyrometer to a PC 27 5 2 Installation 27 5 3 Progra...

Страница 5: ...ment Security note laser beam Indicates to the danger of a built in laser targeting light MB Shortcut for Temperature range in German Messbereich 1 1 2 Terminology The terminology used in this manual...

Страница 6: ...cts caused by LumaSense Technologies The Windows compatible software was thoroughly tested on a wide range of Windows operating systems and in several world languages Nevertheless there is always a po...

Страница 7: ...0176 Email support lumasenseinc com Frankfurt Germany Telephone 49 0 69 97373 0 Email support lumasenseinc com Erstein France Telephone 33 0 3 88 98 98 01 Email support lumasenseinc com 1 6 Shipments...

Страница 8: ...e in humid areas or shipment overseas the device should be placed in welded foil ideally along with silica gel to protect it from humidity The pyrometer is designed for a storage temperature of 20 to...

Страница 9: ...400 C 600 3000 C 300 1700 C 850 2520 C Sub Range Any range adjustable within the temperature range minimum span 51 C Signal Processing Photoelectric current digitized Spectral Range IS 12 TSP 0 94 m a...

Страница 10: ...missivity Response time t90 Clear times tCL of the maximum value storage Temperature sub range Analog output 0 to 20 or 4 to 20 mA Switch points for limit contacts C or F Interface RS232 or RS485 Addr...

Страница 11: ...SP 300 1700 360 Note The calibration adjustment of the instruments was carried out in accordance with VDI VDE directive Temperature measurement in industry Radiation thermometry Calibration of radiati...

Страница 12: ...as temperature or parameters 3 Parameter indicator 4 Parallax free thru lens view finder 5 Setting keys operate with tipp of ball point pen 6 Connection cable 7 Laser warning label only instruments wi...

Страница 13: ...imum value storage via contact see chapter 8 Reference numbers and 12 pin angle connector The cable includes a short RS232 adapter cable with a 9 pin SUB D connector for direct PC communication This a...

Страница 14: ...able at the pyrometer for connection to a processor PC Standard on a PC is the RS232 interface At this interface one pyrometer can be connected if the interface is set to RS232 Only short distances ca...

Страница 15: ...nection of additional analyzing devices Additional analyzing instruments for example a LED digital display instrument only needs to be connected to a power supply and the analog outputs from the pyrom...

Страница 16: ...out temperature measurements The following points must also be observed in the special case of the TSP device 3 4 2 Special aligning notes The size of the spot size is determined first and foremost by...

Страница 17: ...ex works This is the standard setting for TS pyrometers which is achieved with maximum precision at minimum noise values Changing this setting is only necessary if the pyrometer is to be used for othe...

Страница 18: ...17 The aperture is the effective lens diameter It is depending on the objective length The biggest aperture value 17 mm belongs to the fully extended objective S 30 mm the smallest aperture value 13...

Страница 19: ...for each of the 3 focusable optics This corresponds to the longest or the shortest objective length As an example a further value between max and min is shown The objective length S can be measured wi...

Страница 20: ...IS 12 TSP IGA 12 TSP Manual Controls and Installation 20 To ensure consistent document formatting this page was intentionally left blank...

Страница 21: ...2 kBd Interface RS RS232 Limit contacts end of temperature range Hysteresis 2 C Test current output 10 mA off 4 2 Instrument settings All instrument settings can be done directly at the instrument To...

Страница 22: ...lay to the next parameter after the last parameter it changes to the actual temperature reading With the arrow keys and all parameter settings can be displayed Pushing the button longer changes the se...

Страница 23: ...1 1 Steel heavily scaled 0 92 0 93 0 85 to 0 9 Steel rolling skin 0 89 0 88 0 8 to 0 88 Steel molten 0 35 0 3 0 2 to 0 25 Slag 0 85 0 85 0 8 to 0 85 Aluminum bright 0 13 0 15 0 1 Chromium bright 0 32...

Страница 24: ...is used for discontinuous measuring tasks For example objects are transported on a conveyer belt and pass the measuring beam of the pyrometer only for a few seconds Here the maximum value for each obj...

Страница 25: ...cted according to the signal input of the connected instrument controller PLC etc Note Settings for Analog Output 0 to 20 mA or 4 to 20 mA 0 to 20 mA 21 Low or 4 to 20 mA 21 Low setting 21 Low Analog...

Страница 26: ...lly after 1 minute and the instrument is working in the measuring mode 4 14 Wait time tw Using a pyrometer with RS485 it is possible that the connection is not fast enough to receive the pyrometer s a...

Страница 27: ...nc com 5 1 Connecting the pyrometer to a PC The program InfraWin can operate up to two devices Two devices using RS485 may be operated simultaneously by the same interface if two different addresses h...

Страница 28: ...IS 12 TSP IGA 12 TSP Manual Software InfraWin 28 To ensure consistent document formatting this page was intentionally left blank...

Страница 29: ...ptics These optics can be changed against another e g if the lens is scratched or the pyrometer will be used for other measuring distances When changing the optics a recalibration of the pyrometer is...

Страница 30: ...IS 12 TSP IGA 12 TSP Manual Maintenance 30 To ensure consistent document formatting this page was intentionally left blank...

Страница 31: ...ive settings only at setting commands not at enquiry commands Example Entry 00em CR The emissivity setting of the device with the address 00 is returned Answer 0970 CR means Emissivity 0 97 or 97 0 De...

Страница 32: ...00 Baud 4 19200 Baud 7 is not allowed 2 4800 Baud 5 38400 Baud 8 115200 Baud 3 9600 Baud 6 57600 Baud Wait time AAtwXX XX 00 99 decimal Error status AAfs Output XX XX 00 FF 00 no error 01 FF error cod...

Страница 33: ...ce type software version AAve Output XXYYZZ 6 digit decimal XX 07 instruments ID for IS 12 TSP or IGA 12 TSP YY Month of software version ZZ Year of software version Software version in detail AAvs tt...

Страница 34: ...IS 12 TSP IGA 12 TSP Manual Data format UPP Universal Pyrometer Protocol 34 To ensure consistent document formatting this page was intentionally left blank...

Страница 35: ...ngth 25 m 90 connector 3 820 550 Connection cable length 30 m 90 connector 3 820 740 Connection cable length 5 m 90 connector temperature resistant up to 200 C 3 821 120 Additional cable for limit con...

Страница 36: ...IS 12 TSP IGA 12 TSP Manual Reference Numbers 36 To ensure consistent document formatting this page was intentionally left blank...

Страница 37: ...place it see 6 2 6 3 Temperature indication too high Emissivity set too low Set lower correct emissivity corresponding to the material see 4 4 The measurement is influenced by reflections of hot machi...

Страница 38: ...IS 12 TSP IGA 12 TSP Manual Troubleshooting 38 To ensure consistent document formatting this page was intentionally left blank...

Страница 39: ...al 13 Interface 25 L Language 27 Laser targeting light 15 22 Legend 5 Liability 6 M Maintenance 29 P Parameter display 22 Parameter indicator 22 Physical User Interface 12 R Reference numbers Accessor...

Страница 40: ...IS 12 TSP IGA 12 TSP Manual Index 40 W Warm up 13 Warranty 6...

Отзывы: