INFRATHERM 

pyrometer IS 50-LO/GL 

 

 

19

6.5 Accessories 

(option) 

As a standard accessory, the 

INFRATHERM 

IS 50-LO/GL is provided with an air purge tube and mounting bracket (also 

see section 

11

Reference numbers

). For corresponding applications various optional sighting tubes are available. 

 

Air purge unit 

The 

air purge unit with mounting support 

 

protects the lens from contamination with dust 
and moisture. It has to be supplied with dry 
and oil-free pressurized air (approx. 65... 100 
l/min or 4 ... 6 m

3

 /h) and generates an air 

stream shaped like a cone. 
 

 

Ceramic sighting tube 

Mounting support

Bayonet fixing

Inconel-
sighting 
tube 

 
 

6.6 Sighting 

tubes 

 

Stainless steel (Inconel

®

 

Ref. nr.: 3 835 300 
 

An Inconel tube with a 600 mm 
length is available upon request.  

 

 

 

 26.64

 33.5

 
 
Ceramic tube 
 
Ref. nr.: 3 836 830 

 

 27

 35

 
 

7 Optics 

 
A pyrometer operates passively. It receives heat radiation, from the measured object, which passes through the lens and 
is then converted to an electrical signal. The measured object may be at any distance from the pyrometer! However, the 
farther the measured object is from the pyrometer, the larger the area (spot size) that will be measured by the pyrometer. 
The object (or the area on the object) that you want to measure has to be as big (or bigger) than the pyrometer's spot 
size, for that distance.  
The following drawing show the pyrometer's spot sizes (minimum required sizes for the measured object) and the focal 
distances, in relation to the distance from the pyrometer. The spot size at a distance of 0 mm  (a = 0) is 8.8 mm  
(M = 8.8), which is the diameter of the aperture behind the lens. 

58

 

107

 

5000

 

3000

58

31

18

 

2000

 

1200

 

600

 

300

 

8,8

 

0

 

9

 

12

 

10

 

15

 

20

 

12

32

 

Measuring distance 
a [mm]

 

Spot size M [mm]

 

Minimum tube or hole diameter 
(incl. safety tolerances)

 

Optics

 

 

 

 

Note:  

Please note that the optical profiles show nominal dimensions. The spot size diameter or 
the focal distance may be slightly different due to lens tolerances.   

Содержание Impac IS 50-LO/GL

Страница 1: ...IS 50 LO GL INFRATHERM Pyrometer Betriebsanleitung User Manual IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry www imaxsun com...

Страница 2: ...of additional analyzing devices 17 6 Mechanical installation 17 6 1 Overview 17 6 2 Converter 18 6 3 Fibre optical head 18 6 3 1 Ambient temperature 18 6 3 2 Serial number 18 6 4 Minimum bending radiu...

Страница 3: ...C Infrared GmbH have a warranty of two years from the invoice date This warranty covers manufacturing defects and faults which arise during operation only if they are the result of defects caused by I...

Страница 4: ...e must be within the limits of the full temperature range and must have a minimum span of 51 C Emissivity 5 100 set via emissivity control inside the pyrometer Response time 1 s set ex works Accuracy...

Страница 5: ...in a permanent readable condition 4 1 General Each person working with the pyrometer must have read and understood the user manual before operation Also this has to be done if the person already used...

Страница 6: ...n side of the power supply switch board the shield must be open to avoid ground loops 5 1 Connector assignment for the converter Connector 3 pole flange plug series 723 Franz Binder GmbH Co Germany Pi...

Страница 7: ...ble The fibre optic connectors are equipped with a special sealing ring to protect the connection against penetra tion of liquids Attention The light guide end of the fibre optic cable as well as the...

Страница 8: ...on from the measured object which passes through the lens and is then converted to an electrical signal The measured object may be at any distance from the pyrometer However the farther the measured o...

Страница 9: ...erature The emissivity setting of the pyrometer has to be adjusted accordingly Typical emissivity values are listed below Measuring object Emissivity at 0 8 1 1 m molten glass 96 Set up examples at th...

Страница 10: ...e 4 20 mA current output of the pyrometer This is especially important if the temperature control of your system i e oven or furnace temperature is dependent on the output signal from the pyrometer 8...

Страница 11: ...al head 11 2 Accessories 3 820 650 Connection cable length 5 m 3 820 660 Connection cable length 10 m 3 820 670 Connection cable length 15 m 3 820 680 Connection cable length 30 m On request Other len...

Отзывы: