IN 5/9 plus Manual
Data format UPP
•
32
Description
Command
Parameters
Reading temperature
sub range:
AAme
as with mb
Baud rate:
AAbrX
X = 0 to 4
0 = 1200 Baud 2 = 4800 Baud
4 = 19200 Baud
1 = 2400 Baud 3 = 9600 Baud
Reading ambient
temperature:
AAut
Output: stored value, 4-digit hex
e.g. 0258 corresponds to 600 degrees
Entering ambient
temperature:
AAutXXXX
XXXX = value of ambient temperature, 4-digit, hex
XXXX e.g. FFEC corresponds to –20 degrees
- 99dez = FF9Dhex means: automatic, no manual compensation
Marginal values for
ambient
temperature:
AAut?
Output: marginal values for the entry, 2 x 4-digit, hex
e.g. FF9D0384 corresponds to -99 to 900 degrees
Reading marginal
values:
AAmi?
Output: marginal value for entry, always 01
Device address
1)
:
AAgaXX
XX: Device address decimal
00 to 31 variable device addresses
Laser targeting light: AAlaX
X = 0;1 (decimal)
0 = switch off pilot light
1 = switch on pilot light
Internal pyrometer
temperature:
AAgt
Output: XX (dec. 00 to 98, in °C)
Max. internal
temperature:
AAtm
Output: XX (dec. 00 to 98, in °C)
Reading parameters: AApa
Output decimal 11-digit:
Digit 1 and 2 (20...99 or 00): Emissivity
Digit 3 (0 ... 6): Exposure time (t
90
)
Digit 4 (0 ... 8): Clear time max. storage (t
clear
)
Digit 5 (0 ... 1): Analog output
Digit 6 and 7: (00 ... 98): Temperature
Digit 8 and 9 (00 ... 31): Address
Digit 10 (0 ... 4): Baud rate
Digit 11: always 0
Error status:
AAfs
Output 1 Byte hex
Bit 0 = 1: EEPROM error
Bit 1 = 1: Watch dog Reset
Bit 2 = 1: Under voltage reset
Reset
1)
:
Aare
Reset device
Command delay:
AAtwXX
XX = 00 to 20 relative delay value
Reading maximum/
minimum values:
AAmi
Output: 0 or 1
0 = maximum value, 1 = minimum value
Setting maximum/
minimum values:
AAmiX
X = desired setting
0 = maximum value, 1 = minimum value
Serial number:
AAsn
Output: XXXXX (5-digit decimal)
Device model /
software version:
AAve
Output: XXYYZZ (6-digit decimal)
XX = 70 (IN 5/9 plus)
YY = month of the software version
ZZ = year of the software version
Note:
The letter “l” means the lower-case letter of “L”.
1)
After entering these commands, the device carries out an automatic reset. The device needs
approx. 150 ms before it is ready to use and work with the changed settings.
Additional instruction for the RS485 interface:
Requirements to the master system during half-duplex operation:
1.
After an inquiry, the bus should be switched into a transmission time of 3 bits (some older
interfaces are not fast enough for this).
2.
The pyrometer's response will follow 3 ms at latest.
3.
If there is no response, there is a parity or syntax error and the inquiry must be repeated.