IN 5/9 plus Manual
Instrument Settings
•
18
4.1.1
Instrument Switches
See Chapter 5 for the Parameter descriptions / Settings.
4.2
Online/Offline Mode (ONL/OFFL)
4.2.1
Offline (OFFL)
If the settings directly at the pyrometer should be used (emissivity, exposure
time and analog output), the pyrometer must be switched to offline mode
(OFFL). In the offline mode, these parameters can not be adjusted via digital
interface and PC and can only be read. This prevents incorrect set-up or
undesired changes via the interface. All additional available parameters (see
5.2 Settings
only
via serial interface
) can be altered via serial interface.
4.2.2
Online (ONL)
When switched to the online mode (ONL), only the serial interface can be used to set the values
for the parameters and the instrument’s switch settings are ignored. The device is initialized in
“online” mode, i.e. it uses the last values set via serial interface!
4.3
Factory settings
Instrument settings:
Emissivity (
Emi
) = 100%
Exposure time (
t
90
)
= 0.18 s
Analog output (
4/0 mA
) = 0... 20 mA
Online- / offline mode (
ONL/OFFL
) = offline
Interface settings:
Emissivity (
Emi
) = 100%
Exposure time (
t
90
)
= 0.18 s
Analog output (
4/0 mA
) = 0... 20 mA
Baud rate = 19200 Bd
Address = 00
Max./min. value storage = max.
Clear time of max./min. value storage = OFF
Ambient temperature compensation = auto
Wait time t
w
= 02 Bit
Settings:
Emi:
Rotary switch for
Emissivity adjustment
4/0 mA:
DIP switch 2 for
adjusting the analog output
to 0 ... 20 mA or 4 ... 20 mA
ONL/OFFL:
DIP switch 1
for switching
online- / offline mode
t
90
:
DIP switches 3,4,5,6
for adjusting the
exposure time t
90