IN 5/9 plus Manual
Introduction
•
10
Analog Output:
Adjustable 0 ... 20 mA or 4 ... 20 mA (linear)
Load:
Max. 500 Ohm at 24 V (max. 200 Ohm at 18 V)
Digital Interface:
RS232 (RS485 on request)
Parameters:
Adjustable on the pyrometer (in offline mode):
Emissivity, exposure time, 0/4 ... 20 mA analog output range,
online- / offline switch
Readable and adjustable via interface / PC (in online mode):
Emissivity, exposure time, 0/4... 20 mA analog output range, sub
temperature range, automatic clearing of the max./min value
storage, external clearing of the max./min value storage, clear
times of the max/min value storage, address, baud rate, internal
temperature °C / °F, max. / min. selection, activation of ambient
temperature correction
Power Supply:
24 V DC (18 ... 30V DC) nominal, ripple must be less than 0.5 V
Power Consumption:
Max. 70 mA
Isolation:
Power supply, analog outputs and digital interfaces are electrically
isolated from each other
Ambient Temperature: 0 ... 63 °C
Storage Temp.:
-20 ... 70 °C
Rel. Humidity:
Non-condensing conditions
Protection Class:
IP65 (DIN 40050)
Operating Position:
Any
Housing:
Stainless steel
Weight:
410 g
Connection:
12 pin connector
CE-Label:
According to EU directives about electromagnetic immunity
Note:
The calibration / adjustment of the instruments was carried out in accordance
with VDI/VDE directive “Temperature measurement in industry, Radiation
thermometry, Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4.
For additional details on this directive, see http://info.lumasenseinc.com/calibration or
order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany.
2.4
Dimensions
All dimensions in mm